2.
+
Press the SHIFT key to enter the shift state (the SHIFT LED illuminates). When you press the 8
key of the numeric keypad during the shift state, the signal source is switched in the following
order: INT, VCA, SYNC, EXT, ADD, INT. Some signal sources may be skipped and cannot be
selected depending on the AC/DC mode.

3.4.2 Setting the Output Range

You can select the 100 V range or 200 V range. Table 3-10 shows the setting range for each range.
-------- Notes----------------------------------------------------------------------------------------------------------------
 You cannot switch the output range in the output on state.
---------------------------------------------------------------------------------------------------------------------------------
Setting Item
DC voltage
Phase Voltage
1P3W
Line
voltage
3P4W
External Input Gain
Peak current limiter (positive)
Peak current limiter (negative)
RMS current limiter
Voltage Setting Limit
50
Switching the signal source

Table 3-10 The Setting Range per Output Range

100 V range
−220.0 to +220.0
SIN, CLP
0.0 to 155.0
ARB
0.0 to 440.0
SIN
0.0 to 310.0
SIN
0.0 to 268.4
0.0 to 220.0
Setting Range
200 V range
−440.0 to +440.0
0.0 to 310.0
0.0 to 880.0
0.0 to 620.0
0.0 to 536.8
0.0 to 440.0
Refer to 10.13
Refer to 10.14
PROGRAMMABLE AC POWER SOURCE
Unit
V
Vrms
Vp-p
Vrms
Vrms
Times