10. Specifications

10.22 External Signal Input

The external signal input works differently depending on the selection of the signal source.

10.22.1 External Synchronous Signal Input (Signal Source SYNC only)

This is the signal to synchronize the frequency of the internal signal source with the one of the
external signal source.
When the signal source is SYNC, you can select whether to synchronize with this external signal
input or the power input frequency of the product. When synchronizing with the power input
frequency, no signal input is needed.
Synchronization signal
source switch
Synchronization
frequency range
Input terminal
Input impedance
Threshold of input
voltage
Minimum pulse width
Nondestructive
maximum input voltage

10.22.2 Voltage Setting Signal Input (Signal Source VCA only)

This is the signal to set the output voltage amplitude of the internal signal source (DC input).
Output voltage (Vop) = Voltage setting signal (Vdc) x Gain (Vop/Vdc)
Example 1) For the AC mode, signal source = INT, the gain of 100.0, and the voltage setting
signal input of 1 Vdc, the output voltage is 100 Vop
Example 2) For the AC mode, signal source = INT, the gain of 141.4, and the voltage setting
signal input of 1 Vdc, the output voltage is 141.4 Vop(=100 Vrms)
Gain setting range
Setting resolution
Gain accuracy
Input terminal
Input impedance
Input voltage range
Nondestructive
maximum input voltage
Note: In the polyphase model, polyphase system, and polyphase output of the Multi-phase model,
the setting is common to all the phases.
262
All models
External synchronization signal (EXT) or
Power input (LINE)
BNC connector (rear panel, unbalanced)
Specification
100 V range: 0.0 to 220.0 times
200 V range: 0.0 to 440.0 times
(DC, 45 Hz to 65 Hz, the gain is the factory default, the rated output
BNC connector (rear panel, unbalanced)
Also used as the external synchronization signal input
±2.2 V (A/D resolution 10 bit)
40 to 500 Hz
1 MΩ
TTL level
500 μs
±10 V
0.1
±5%
voltage, no load)
1 MΩ
±10 V
PROGRAMMABLE AC POWER SOURCE
Factory default
LINE
Factory default
100
200