Service functions can be used to select institutional defaults and to access
information about the patient or instrument. Only a Nellcor Customer Service
Engineer should access some of the items available through the service
functions. These items will be noted in the text that follows.
All service functions are accessible only if the DOC-10 pulse oximetry cables
are disconnected from the monitor.
1. Turn on the N-595 by pressing the POWER ON/OFF button.
2. Wait for monitor power-on self-test to complete.
3. Simultaneously press and hold the LIGHT softkey and the CONTRAST
button until the service softkeys are displayed.
Power-On Settings and Service Functions
Setting
ASCII
Midrange
Pleth
SpO
2
English
Normally Low
4 of 10
40 beats per minute
170 beats per minute
Off
Yes
%SpO
2
31