The SpaceLabs monitor provides both audible and visual alarm indications
for equipment interfaced through the Universal FlexPort
Silencing the N-595 alarms will also silence the SpaceLabs monitor alarms.
The monitors must be able to sound an audible alarm in order to maintain
patient safety.
Note:
The parameters setup for the SpaceLabs bedside monitor interface
may be saved as institutional default settings. See Setting Industrial
Defaults (Sample) page 40. Use care when performing this procedure
because all settings will be saved as institutional default settings.
The N-595 sends SpO
, pulse rate, and alarm status data to the Marquette
2
monitor.
The Marquette monitor requires an Octanet
with the N-595 pulse oximeter. The interface module comes with an interface
cable, GE Marquette part number 417961-033, that connects to the Nellcor
interface cable.
The RS-232 hardwire interface cable has a DB-15 connector for the
N-595 and the applicable connector for the Marquette Octanet
module cable. Nellcor cable part number 902254 is recommended for this
interface.
Corrupt data will be indicated by a Communications Error displayed on the
Marquette monitor.
When the N-595 is in the Marquette mode of operation the interface baud rate
is automatically set to 9,600 bits per second.
The GE Marquette monitor only sounds audible alarms for equipment
interfaced through the Octanet
N-595 audible alarm has no effect on the GE Marquette monitor sounding an
alarm.
Note:
The parameters setup for the Marquette bedside monitor interface
may be saved as institutional default settings. See Setting Industrial
Defaults (Sample) page 40. Use care when performing this procedure
because all settings will be saved as institutional default settings.
The Datex-Ohmeda monitor AS/3 must be configured for communications
with the Nellcor N-200 monitor in order to communicate with the N-595
monitor. Refer to the AS/3 operator's manual for instructions on configuring
the AS/3 monitor.
The N-595 sends SpO
, pulse rate, and alarm status data to the Datex AS3
2
monitor.
TM
interface module to interface
TM
interface module. Silencing the
TM
interface module.
TM
interface