Agilent (Hp) Communications; Spacelabs Communications - Nellcor OXIMAX N-595 Service Manual

Hide thumbs Also See for OXIMAX N-595:
Table of Contents

Advertisement

Agilent (HP)
Communications
SpaceLabs
Communications
N-595
The N-595 sends SpO
, pulse rate, and alarm status data to the Agilent
2
monitor.
The Agilent monitor requires an Agilent VueLink
module (A05 option) to interface with the N-595 pulse oximeter.
The RS-232 hardwire interface cable has a DB-15 connector for the
N-595 and the applicable connector for the Agilent monitor. Nellcor cable
part number 902256 is recommended for this interface.
A blank screen on the Agilent monitor will indicate corrupt data. The Agilent
monitor will detect corrupt data in less than 100 milliseconds.
When the N-595 is in the Agilent mode of operation the interface baud rate is
automatically set to 19,200 bits per second.
WARNING: Do not silence the N-595 audible alarm or decrease its
volume if patient safety could be compromised.
The Agilent monitor only displays visual alarm indications for equipment
interfaced through the Agilent VueLink
N-595 monitor must be able to sound an audible alarm in order to maintain
patient safety.
Note:
The parameters setup for the Agilent bedside monitor interface may
be saved as institutional default settings. See Setting Industrial
Defaults page 40. Use care when performing this procedure because
all settings will be saved as institutional default settings.
The N-595 sends SpO
, pulse rate, and alarm status data to the SpaceLabs
2
monitor.
The SpaceLabs monitor requires a Universal FlexPort
interface with the N-595 pulse oximeter.
Corrupt data will be indicated by a Communications Error displayed on the
SpaceLabs monitor.
When the N-595 is in the SpaceLabs mode of operation the interface baud rate
is automatically set to 9,600 bits per second.
WARNING: Do not silence the N-595 audible alarm or decrease its
volume if patient safety could be compromised.
Data Port Interface Protocol
TM
Aux Plus B interface
TM
Aux Plus B interface module. The
TM
interface module to
95