Setting
PROJECTION SCALE
Enter individual
PPM
EDM SIGNAL
REFLECTION
4.3
Description
Access
The refraction correction is taken into account in the calculation of the height
differences and the horizontal distance. Refer to "14.7 Scale Correction" for the
application of the values entered in this screen.
When PPM=0 is selected, the Leica standard atmosphere of 1013.25 mbar,
12°C, and 60% relative humidity will be applied.
This screen enables entry of the scale of projection. Coordinates are corrected with
the PPM parameter. Refer to "14.7 Scale Correction" for the application of the values
entered in this screen.
This screen enables the entry of individual scaling factors. Coordinates and distance
measurements are corrected with the PPM parameter. Refer to "14.7 Scale
Correction" for the application of the values entered in this screen.
This screen tests the EDM signal strength (reflection strength) in steps of 1%. Enables
optimal aiming at distant, barely visible, targets. A percentage bar and a beeping
sound, indicate the reflection strength. The faster the beep the stronger the reflection.

Communication Parameters

For data transfer the communication parameters of the instrument must be set.
1. Select Settings from the MAIN MENU.
2. Select Comm from the SETTINGS MENU.