Setting
Field
Description
ALL
DIST
USER Key 1 /
Configures
Key 2
Functions".
Tilt Corr.
Off
1-axis
2-axis
If the instrument is used on an unstable base, for example a shaking
platform or ship, the compensator should be deactivated. This
avoids the compensator drifting out of it's measuring range and
interrupting the measuring process by indicating an error.
Sets the trigger key with the same function as ALL.
Sets the trigger key with the same function as
DIST.
or
with a function from the FNC menu. Refer to "6
Tilting compensation deactivated.
Vertical angles refer to the plummet line.
Vertical angles refer to the plummet line and the
horizontal directions are corrected by the standing
axis tilt.
For corrections depending on the Hz Corr: setting,
refer to the table "Tilt and horizontal corrections".
FlexLine, 45