10.9.2
Error messages
Table 15: Error table: Malfunction message
Initiator
Text
EEPROM
Spectro com.
Zero com.
Temp control
com.
Visor com.
Filter com.
Mirror com.
Lamp com.
Visor fault
Visor values
Visor no signal
System
Lamp fault
Mirror adj. End
Zero adj. mc
adj.
Spectro para.
Purge air sig‐
nal
Temp control
out of range
Extinction calc
Reference calc Failure
IIR Filter
Interpolation
Eval modul
com.
File conditions
File espec
File cact
File measval
8015558/1O6S/V1-9/2024-08 | SICK
Subject to change without notice
Classifica‐
Description
tion
EEPROM parameters corrupted or not compati‐
ble after software upgrade.
Failure
Communication fault with spectrometer.
Failure
Communication fault with zero point reflector.
Failure
Communication fault with temperature control
Failure
unit.
Communication fault with visor module.
Failure
Communication fault with control filter ele‐
Failure
ment.
Communication fault with mirror tracking.
Failure
Communication fault with lamp electronics.
Failure
Failure
Error of visor signals. Signal distorted or zero.
Visor signals outside the valid range.
Failure
Failure
All 4Q signals below threshold parameter.
Lamp does not go on.
Failure
Mirror tracking has reached maximum posi‐
Failure
tion.
Beam tracking during adjustment not possible.
Failure
Failure
No correct parameters saved in spectrometer.
Digital input signals purge air error.
Failure
Temperature regulation measurement outside
Failure
the valid range.
Failure
Error in extinction calculation.
Error in reference calculation.
Failure
Error during IIR filtering.
Failure
Error in interpolation calculation.
Error in communication with software evalua‐
Failure
tion module.
Failure
Error during condition file access.
Failure
Error during extinction file access.
Failure
Error during lambda coefficient file access.
Failure
Error during measured value file access.
TROUBLESHOOTING
Possible causes/remedy
•
Software upgrade: Reset parameters. Load
saved parameters
•
Defect: Reload backup. Possibly replace
hardware.
Error in I2C connection with spectrometer. Plug
connector / check for hardware defect.
Error in I2C connection with zero point reflector.
Plug connector / check for hardware defect.
Error in I2C connection. Plug connector / check
for hardware defect.
Error in I2C connection with visor module. Plug
connector / check for hardware defect.
Error in I2C connection with control filter ele‐
ment. Plug connector / check for hardware
defect
Error in I2C connection with mirror tracking. Plug
connector / check for hardware defect.
Error in I2C connection with lamp electronics.
Plug connector / check for hardware defect.
Check signals and parameters.
Hardware defect. Electronics not adjustable
(amplification too high).
Check alignment, reflector, contamination.
Lamp is defective. Replace lamp,
the sender lamp", page
67.
Check alignment,
see "Alignment check (option)
- automatic optical alignment check", page
Please contact SICK Customer Service.
Check purge air supply.
Excess temperature switch-off active for temper‐
atures > 70 °C. Switches back on automatically
when < 65 °C.
Please contact SICK Customer Service.