10
TROUBLESHOOTING
Error
All measured values blink
10.9
Error messages
10.9.1
Example of an error message
74
O P E R A T I N G I N S T R U C T I O N S | GM32 Ex
Possible causes
•
The lever for the diaphragm is set to
"closed".
The error message displayed here is shown on the device display of the in-situ gas
analyzer GM32 Ex. Refer to the SOPAS Manual for the display via SOPAS ET.
Measuring
Failure
1 / 1
04/29 01:57
1
System:
3
Mirror adj. end
4
/Menü/Diag/Failure
Back
Figure 40: Error message display
Message number / number of messages
1
Date (dd/mm) time (hh:mm:ss)
2
Initiator
3
Error message
4
NOTE
All error messages are listed in a Table, see Chapter Error Messages.
The definition of "Initiator" is shown at the end of the Table.
Remedy
Set lever to "open".
►
1.1.1
POWER
2
FAILURE
MAINTENANCE
REQUEST
8015558/1O6S/V1-9/2024-08 | SICK
Subject to change without notice
Need help?
Do you have a question about the GM32 Ex and is the answer not in the manual?