1. 3 Designated users The GM32 may be operated by competent persons only who, based on their device-specific training and knowledge of the device as well as knowledge of the relevant regulations, can assess the tasks given and recognize the dangers involved.
Product description 2. 2 The GM32 gas analyzer serves for continuous measurement of the gas concentrations in industrial plants. GM32 is an in-situ measuring system. Measurement is performed directly in the gas carry- ing duct. Components: SO , NO, NO...
SOPAS ET (PC program) 2 . 3 SOPAS ET can be used to set the GM32 parameters and provides access to the GM32 log- book. SOPAS ET runs on an external PC connected via the Ethernet interface (→ p. 22, Figure 4) to GM32.
The basis for the determination of the sampling point is the preceding proj- ● ect planning (e.g. based on the SICK application questionnaire), the infor- mation in the final inspection record of the GM32 and the regulations of the local authorities. The equipment operator is responsible for the determination of the sam- ●...
GM32 is protected from excess temperatures (→ p. 67, §9.2). Installing the connection unit 3. 3 The length of the line to the GM32 complies with project planning. ● Provide threaded bolts (4) to screw on the connection unit (→ p. 72, §31) and screw on the connection unit.
Preparation on the Gas Duct Side Installing the purge air unit (for GMP probe) 3 . 4 Maximum length of line to GM32 complies with project planning. ● IMPORTANT: Adequate purge air pressure Ensure the purge air pressure is adequate to push the purge air into the gas duct.
Only let the work described in the following be carried out by skilled electri- cians familiar with potential risks. IMPORTANT: Before signal connections are established (also with plug connections): Disconnect the GM32 and any connected devices from the electrical sup- ply. Otherwise the internal electronics can be damaged. Connecting I/O interfaces (option) 3.5.2...
Power supply 100 .. 240 V / 50 .. 60 Hz The power supply must remain switched off until the GM32 is to be put into operation. 3 For GMP probe: Lay the electric lines to the purge air unit.
Access to the logbook is performed exclusively via SOPAS ET → p. 13, §2.3. The GM32 signals a malfunction or an unsafe operational state with status sig- nals (option) (→ electrical connection diagram). List of error messages and measures for clearing malfunctions →...
See SOPAS ET in the “Digital outputs” menu. Alignment check (automatic optical alignment) (option) 5.2.5.2 The optical alignment of GM32 is automatically adjusted in this menu. Perform this work only when the SR-unit is at operating temperature (in operation for at least 30 minutes).
Value X Value Y Manual optical alignment of the GM32. 1 Press the “Start” button: The GM32 goes to a defined state. A crosshair with a focal point and X/Y values is shown on the screen. 2 Tolerances: X: –0.05 ... +0.05 Y: –0.05 ...
3 Protect the openings of the SR-unit and measuring probe from atmospheric influences (preferably with the transport safety devices, according to → p. 27, §4.4). 4 Pack the GM32 for storage or transport (preferably in the original packing). 5 Store the GM32 in a dry, clean room.
Maintenance Preparation work 7 . 2 IMPORTANT: Some tasks cause GM32 to switch to malfunction Activate the Maintenance mode before starting the work (→ p. 44). IMPORTANT: Do not switch off the purge air ⊗ Do not switch the purge air unit off as long as the SR-unit is still on the gas duct.
Replacing the sender lamp 7 . 5 1 Switch off the GM32 using the fuse on the operator side. 2 Loosen 5 screws on the rear side of the SR-unit and swivel out the rear side. 3 Pull out the lamp cover.
Do not disconnect the protective conductor connections inside or outside the device. IMPORTANT: Damage by voltage Before signal connections are established (also with plug connections): Disconnect the GM32 and any connected devices from the electrical sup- ply. Otherwise the internal electronics can be damaged. Measured value blinks 8.
On systems with pressure difference sensor: An appropriate error message occurs. ● Rise of housing temperature. ● Unusually rapid contamination of the window of the GM32. ● Check the purge air unit Remove the purge air hose on the SR-unit: A strong air flow must be noticeable.
Page 68
Check cycle for zero and reference point, equivalent to QAL3 (option) ● per module For more information and technical information on the GM32 system and its com- ponents, see the following documents: Data Sheet GM32, Cross-Duct Version, Part No. 8012710 ●...
Page 76
You will find our local subsidiary or agency at: www.sick.com Your local sales and service partner SICK MAIHAK GmbH | Germany Nimburger Str. 11 | 79276 Reute | www.sick.com Phone +49 7641 469-0 | Fax +49 7641 469-11 49 | info.pa@sick.de...