3
PRODUCT DESCRIPTION
1
2
3
5
Figure 5: Measuring principle of check cycle
Output of measured values during the check cycle
Signal output during the check cycle (optional digital output or
OPC interface)
Output of determined zero and reference values on analog
outputs
Signal output during the output
Display of zero and reference values, the last check cycle as
well as QAL3 values in SOPAS ET
Display when check of NO cell failed
18
O P E R A T I N G I N S T R U C T I O N S | GM32 Ex
6
7
4
Last valid measured value
Not_measuring
•
•
Output_control values (optional digital output or OPC inter‐
face)
Output sequence:
1 Zero values for 90 s
2 Reference values for 90 s
Menu: Diagnosis/check values
•
•
•
•
8
Measuring
1
Zero point
2
Check
3
Data recorder
4
Swivel element
5
Zero point reflector
6
Zero point
7
Check point (70% fsc)
8
Directly after the check cycle
On request via a digital input (option)
All interfaces: Results of NO cell
All interfaces: "0" is output instead of zero and reference
values
Analog output: Live Zero
Results of zero and reference measurement are irrelevant
8015558/1O6S/V1-9/2024-08 | SICK
Subject to change without notice