Page 1
Title Page MMMOPERATING INSTRUCTIONS MMMI O P E R A T I N G I N S T R U C T I O N S GM901 Carbon Monoxide Gas Analyzer, Cross-Duct Version...
Page 2
Original document This document is an original document of SICK AG. O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
PROFIBUS interface (if installed) ..........19 Optional accessories..................20 Transport and storage................21 Storage ......................21 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Page 4
8.1.2 Display ..................46 8.1.3 Status LEDs .................. 46 O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
Page 5
Troubleshooting on receiver............80 11.3.3 Troubleshooting on the control unit..........81 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Page 6
14.1 Spare parts ....................91 14.2 Options, accessories ..................91 Conformities ..................... 92 O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
– Final inspection protocol NOTE: Observe all documents provided. 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
SICK AG always assumes the integrity and confidentiality of data and rights affected in connection with the use of the products are ensured by the customer. O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK...
Page 9
8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
– Available for reference. – Passed on to new owners. O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
The person setting the system up is responsible for the safety of the system in which the device is integrated. 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Certain configurations of the GM901 measuring system demand immediate or short term measures to protect the measuring system should the purge air supply fail. O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK...
[1] See Section 9.4.8 [2] Based on 500 ppm CO, the uncertainty is lower for higher concentrations O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK...
Page 15
[2] Based on 500 ppm CO, the uncertainty is lower for higher concentrations 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Page 16
25 vol. % 370 °C 350 °C 330 °C O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
Applications up to 250 °C Applications up to 430 °C 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
1 analog input 1 analog output Power supply: 115/230 V AC O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
Measured variable mg/m (s.c.) mg/m (a.c.) 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Temperature sensor PT 100 ● Weather protection hood ● O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
Store all components of the measuring device in a dry, clean area. 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
▸ Check all components for externally perfect delivery condition. O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
DIN174 (2”) Steel tube 50 x 5 DIN 2391 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Fig. 5: Aligning the flange using an optical adjustment device Adjustment tube Light source O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
(definition, see Fig. 6). Keep the measures available for commissioning. 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Pull the rubber band over this connection gap. 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Quick-release clamp Nuts to adjust the optical axis O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
Remove the optical adjustment device and fasten the sender and receiver using the quick-release clamps. 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Insert the side holding catches into the counterpieces, twist and lock in. O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK...
Screw the control unit tight to the four fastening brackets. 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
PE conductor for the connection on the outside of the control unit (to comply with the ● EMC regulations) O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
▸ Connect the system components as shown in Fig. 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
For information on the purge air unit, see the Operating Instructions of the purge air unit (SLV4). O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK...
The control unit is delivered from the factory as 230 V AC version. 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
DO3 Limit 2 The control unit is delivered from the factory as 230 V AC version. O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK...
Standard cable (2 m) +24 V CAN-H CAN-L CAN-GND 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Check (measure) the active measuring distance, see “Measuring distance definition”, page 25 O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
Enter SPAN value, see “Changing the calibration values”, page 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Reset temperature parameters to reflect process conditions, see “Temperature”, page O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
Press CAL on the operating panel of the control unit to start the zero adjust (see “Performing zero adjust”, page 69). 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
▸ Manual SPAN test (optional) for linearity control. 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Factory data assigned to the GM901 receiver. (Individual for each device) O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
“Maintenance”, page 71 “meas”: Measurement see “Measuring mode”, page 47 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Read the malfunction message, see Diagnosis mode (diag) ● O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
T=150 °C measuring mode Measuring T=150°C 1128 Nm 3 2000 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Signal too high Signal too high back back back back O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
(see “Malfunctions”, page ▸ Press “Arrow left” (back) 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
CC: 500.0 AG: 12.04 AG: 12.04 back back back back O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
Submenu “Service”, see “Service”, page 68 Calibration values, device-specific 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Press “Arrow left” (back) edit: Enter edit: Enter back back O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
Press “Arrow left” (back) back back edit: Enter edit: Enter 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Press “Arrow left” (back) back back back back edit: Enter edit: Enter O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
Press “Arrow left” (back) back back edit: Enter edit: Enter 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Press “Arrow left” (back) back back edit: Enter edit: Enter O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
Active : 2000 mm back back edit: Enter edit: Enter 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
0 °C Scale High 250 °C Select back back O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
Page 59
0 °C Scale High 250 °C back back Edit: Enter 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Page 60
Press “Enter” Scale High 250 °C back back select O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
Page 61
Input High 20.0 mA 20.0 mA back back select 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Substitute : 00.0 % “Arrow up” ▸ Press “Enter” back back select O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
Substitute : 1013 hPa back back edit: Enter edit: Enter 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Enter edit: Enter edit: Enter edit: Enter O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
Zero : Zero : back back edit: Enter edit: Enter 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Press “Enter” Software Revision Software Revision Configuration back back O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
Page 67
Press “Arrow left” (back) Software Revision Software Revision Configuration back back 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Zero Adjust: Zero adjust ● SPAN Test: Manual SPAN test ● O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
: +0,0... : +0,0... back Save: Enter Save: Enter 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Use “Back” to terminate span adjustment. O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
Measuring T=150 °C Measuring T=150°C 1128 Nm 3 2000 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Test Analog Out Test Analog Out Test Relay Test Relay back O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
Analog-In can be tested using the displayed sample gas temperature. 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Relay 2: Off Relay 2: Off back edit: Enter edit: Enter O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
The device master file (GSD) can now be configured via the PROFIBUS Master for operation of the GM901. 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Yearly Drift check (zero point/adjust and sensitivity check with test cells) O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
(see “Changing the measuring range”, page 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Motor in the receiver unit defective Exchange receiver unit ● O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
LED: On when the lamp is on and the chopper disk rotates Adjustment screw: 4 x Sender housing 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Cable to control unit Adjustment screw: 4 x Receiver housing O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
If no error is found, connect the system components one after each other. 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Page 82
If the sensor values of the GM901 are outside these value ranges, please contact SICK Customer Service for remote diagnosis. O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK...
– Electronics: Capacitors, rechargeable batteries, batteries. – Display: Liquid of LC display. 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Power supply Voltage: 24 V DC Supply via control unit O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
Frequency: 50 Hz / 60 Hz Power input: ≤ 50 VA 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Built-in components Integrated 24 V power supply unit for supply of the sender/receiver unit O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK...
240,7 / 404,7 Ø 39,5 298 / 462 90° 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Dimensional drawing, control unit Fig. 28: Control unit AWE (sheet steel enclosure version), dimensions in mm O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK...
Fig. 30: Flange with tube, DN125 (dimensions in mm) Ø 18 240 / 500 / 1.000 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Fig. 31: Weather protection hood for sender/receiver unit (dimensions in mm) 282,5 12,5 O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
SPAN test cell CO 4,000 ppm 2127627 SPAN test cell CO 10,000 ppm 2127628 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
EN 61326, Electrical equipment for measurement, control and laboratory use - EMC ● requirements O P E R A T I N G I N S T R U C T I O N S | GM901 8008250/1F7Y/V4-0/2022-06 | SICK Subject to change without notice...
Page 93
CONFORMITIES 8008250/1F7Y/V4-0/2022-06 | SICK O P E R A T I N G I N S T R U C T I O N S | GM901 Subject to change without notice...
Page 94
Phone +7 495 283 09 90 E-Mail info@sick.com E-Mail office@sick.com.gr E-Mail info@sick.ru Vietnam Hong Kong Singapore Phone +65 6744 3732 Phone +852 2153 6300 Phone +65 6744 3732 E-Mail sales.gsg@sick.com E-Mail ghk@sick.com.hk E-Mail sales.gsg@sick.com Detailed addresses and further locations at www.sick.com...