Page 1
O P E R A T I N G I N S T R U C T I O N S GM32 Ex In-situ Gas Analyzer Cross-Duct Version Installation, Operation, Maintenance...
Page 2
Original document This document is an original document of SICK AG. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
Scope of delivery..................5.3.1 Checking the delivery state............. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Page 4
8.2.3 Setting the display contrast............ 8.2.4 Menu tree................. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
Page 5
12.5 Purge air fixture: Reflector unit GM32 Ex..........8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Page 6
13.2 Electrical protection.................. 13.3 Ex certifications..................Index....................100 O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
The focus here is on product availability and features. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Notice Hazard which could result in property damage. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
Important technical information for this product Important information on electric or electronic functions 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Wear suitable protective goggles and protective gloves when work on the open device with the power supply connected is performed. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK...
Page 11
The person setting the system up is responsible for the safety of the system in which the device is integrated. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Figure 2: GM32 Ex connection unit Intended use O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
After switching off the main power supply: Wait 20 minutes before opening the enclosure. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Must be available for reference. Must be passed on to new owners. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
Figure 3: Type code for GM32 Ex Ex version designation 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Weatherproof cover SOPAS ET (PC program) SOPAS ET enables: O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
Signaling of maintenance request: Deviation from setpoint value > ± 2% of the FS. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Page 18
Results of zero and reference measurement are irrele‐ vant O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
Evaluation unit Overpressure control p/T sensor p sensor O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
The operating company is responsible for the evaluation of the alarm signal. See the Pressurized Enclosure System Manual. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Maintenance Chapter, see "Check/replace the desiccant cartridge", page 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
IEC 60079-19: Equipment repair, overhaul and reclamation • Local work safety regulations O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
Check the scope of delivery according to the order confirmation/delivery note. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
When the measuring channel is hot, plan the duct and flange insulation so that the device is protected against high temperatures. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK...
Page 27
Install the adjustment device (light source on SR-unit side, adjustment tube on reflector side) onto the flanges. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
"Dimension drawings: Connection unit version 3G", page O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
Attach clearly visible warnings against accidental switching-off to all switching devi‐ ces where the purge air supply can be switched off. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK...
Page 31
Switch the power supply of the purge air unit on again. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
This is also valid for plug connections. Disconnect the GM32 Ex and any connected devices from the voltage supply. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK...
Page 33
Install all lines in accordance with EN61010-1 and EN60079-14. Protect lines against electrostatic charges. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Connection line, purge air monitor, pressure switch Connection line, purge valve SVP3 G3G 8"-300L EEx p O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK...
Never lay power supply cables directly next to signal cables. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Figure 15: Connection diagram of connection unit for variant 3G / Ex zone 2 O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK...
Page 37
Max. outer capacity Max. outer inductivity L/R ratio (inductivity/resistance) 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
SOPAS. For more information, see the SOPAS Operating Instructions. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK...
It is under overpressure. Protect the connecting hose from possible damage. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Connection of potential equalization Screw the grounding conductor tight on the screw terminal. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK...
Check the gland screw fittings, sealing plugs and flanges for leak tightness and tight seat. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
It is under overpressure. Protect the connecting hose from possible damage. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK...
Connect the lines in the connection unit. Screw M screw fittings tight. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
) of the equipment ground tight to the screw terminal. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
"Optical fine alignment of the SR-unit ", page 52 O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
"Swiveling out and removing the SR-unit or reflector unit", page 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Page 48
Fit the sealing ring over the gap. Push the two hose clamps over the sealing ring and screw tight. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK...
Insert the adjustment device with tube into the device flange. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Connecting the protective gas on the SR-unit The protective gas connection is on the intermediate enclosure. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK...
Switch the main power supply of the GM32 Ex on (at an external source). 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
"Alignment check (option) - automatic optical alignment check", page O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
Mounting ring Bottom mounting ring Locking bolts Installation plate 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Page 54
Position the cover on the installation plate from the top. Engage the locking bolts and close again. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK...
On the GM32 "Pro" device version, the operator panel is on the right side of the SR-unit housing. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
At least one error code is active. MAINTENANCE REQUEST O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
Move cursor to the left. Move cursor to the right. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
1.4.1 IP IP address 1.4.2 M Subnet mask O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
The results of the last check cycle can be called up in this menu item. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Only carry out this check when the SR unit is at operating temperature (minimum 30 minutes in operation). O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK...
Page 61
Perform adjustments slowly and wait for approx. 20 seconds until the display on the monitor has been updated. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Maintenance Maintenance mode activ ******* MAINTENANCE REQUEST Back O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
Page 63
Exit Maintenance mode, current measured value display: Press the Meas button. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Wait 20 minutes after switching off the main power supply before opening the enclosure. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK...
Page 65
8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Consumable parts, expendable parts and spare parts Recommendation for 2 years operation O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
Carry out all work skillfully and in accordance with the delivered documentation. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Swivel the SR-unit out. Quick-release fasteners Hinge pin O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
Check flanges and screw fittings for firm seating. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
2010549 The desiccant cartridges are in the sender/receiver unit next to the window. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK...
The sender/receiver housing may be opened only in a dry, clean environment. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
This is also valid for plug connections. Disconnect the GM32 Ex and any connected devices from the voltage supply. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK...
Page 73
Screws of the sender lamp Figure 39: Replacing the sender lamp 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Page 74
13 Screw the rear cover tight. NOTE Adjustment is not required. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
Disconnect the GM32 Ex and any connected devices from the voltage supply. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Disconnect the device from the main power supply and signal voltage. Prevent unauthorized or unintentional start-up. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK...
Failure of the control unit triggers the display of an alarm signal. There is no automatic switch-off. NOTE See Pressurized Enclosure System Manual. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
GM32 Ex. Refer to the SOPAS Manual for the display via SOPAS ET. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK...
Page 79
The definition of “Initiator” is shown at the end of the Table. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
File measval Failure Error during measured value file access. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
Page 81
Incorrect configuration of I/O module or I/O Maint. Error module defective. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Page 82
Maint. Maint. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
Page 83
This Table also contains recommended solutions that can only be performed by specially trained personnel. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Check the power supply of the connection unit. Contact SICK Customer Service when power supply present. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK...
Wait 20 minutes after switching off the main power supply before opening the enclosure. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Page 86
Switch off the purge air unit only after removal of the SR-unit from the gas duct. O P E R A T I N G I N S T R U C T I O N S | GM32 Ex...
Observe relevant local conditions for disposal of industrial waste. 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Page 88
Electronics: Capacitors, rechargeable batteries, batteries • Display: Liquid of LC-Display O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
27th BImSchV • EN 14181 • EN 15267-3 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
≤ 550 °C , versions for higher temperatures on request Process pressure –60 hPa ... 200 hPa O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
679 mm x 630 mm x 158 mm Weight ≤ 16 kg 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
The housing of the sender/receiver unit can be swiveled to the left or right of the device flange (max. 180°/105°). O P E R A T I N G I N S T R U C T I O N S | GM32 Ex...
8.0 ... 10.0 2046854 10.0 ... 12.0 2046858 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Hose connection for 8 mm diameter • Outlet Thread G 1" O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK Subject to change without notice...
2 mbar (0.03 psi) (relative) Signal pressure 1.5 mbar (0.022 psi) (relative) 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
X II 3G Ex pzc op is [ia] IIC T3 Gc IECEx GM32 Ex 3G O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK...
Page 99
If the duct has underpressure against the atmosphere, this area may correspond to Zone 2 8015594/YV18/V1-8/2017-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM32 Ex Subject to change without notice...
Potential equalization on the purge air fixture....................... 40 Power connection in connection unit........................44 O P E R A T I N G I N S T R U C T I O N S | GM32 Ex 8015594/YV18/V1-8/2017-08 | SICK...