Recover From A Power Failure - Micromeritics TRISTAR II PLUS Operator's Manual

Surface area and porosity analyzer
Hide thumbs Also See for TRISTAR II PLUS:
Table of Contents

Advertisement

11 Troubleshooting
If the O-ring remains inside the sample port, use the O-ring removal tool from the
accessory kit [tool part number 004-54618-00].
5. After the new O-ring is in place, insert the sample tube back into the sample port until it is
fully seated.
6. Slide the sample tube connector nut up the tube until it comes in contact with the port fitting
(the ferrule and O-ring will move along with the connector nut). Then, turning clockwise,
hand tighten the connector nut to the sample connector.
R
ECOVER FROM A
The analyzer saves entered and collected data in case of power failure. File parameters and any
other data entered will still be present when power is restored. If an analysis was in progress when
the power failure occurred, it will be canceled when the analyzer restarts. Any data collected
during the analysis will still be present, but the analysis should be restarted in order to produce
complete results.
11 - 20
P
F
OWER
AILURE
TriStar II Plus Operator Manual
303-42800-01 (Rev M ) — Sep 2023