The Langmuir calculation determines the surface area of a sample by relating the surface area to
the volume of gas adsorbed as a monolayer. Langmuir uses a single layer model.

Langmuir Report

Selections
Pressures [button]
TriStar II Plus Operator Manual
303-42800-01 (Rev M ) — Sep 2023
Description
This option is available when the sample file has a status of
Analyzing or Complete. Use to enter a range of pressure points to be
included in the report or to modify table values for pressure points.
Calculation pressure range. Enter the minimum and maximum
pressures to be used in the pressure table. To exclude a point from
the calculations used to generate the report, select Exclude.