3 Maintenance and Troubleshooting
R
F
ECOVER
ROM A
If an analysis was in progress when a power failure occurred, completed cycles will be saved. The
instrument is vented on startup to ensure that it is in a safe state.
Even though the instrument saves data during a power failure, any operation should
be restarted to ensure complete results.
P
I
OWER
NSTRUMENT
This instrument does not have a power switch and must be disconnected from the
wall outlet when powering off; therefore it is important to position the instrument where
the wall outlet is easily accessible.
When the analyzer is powered on, after a few seconds, the system vents automatically. Allow
approximately 30 minutes for temperature stabilization before performing analyses. For analyses
that require very precise results, allow the analyzer to warm a minimum of two hours.
Wait a minimum of 30 seconds between powering off and powering on the instrument.