11 Troubleshooting
P
L
ERFORM A
EAK
Unit [n] > Diagnostics > Start Diagnostic Test
A Micromeritics Service Representative may request that a leak test be performed to determine if
there is a system leak and may also require a copy of the report generated by this test.
The test provides:
Prompts on preparing the analyzer for the test.
n
Approximate time period of the test.
n
Prompts in which an operator response is required.
n
1. Select the test to run.
2. Select Report After Test and choose Preview as the destination.
3. Click Start.
4. Verify all tests have a Passed status by selecting the tabs and looking for the Passed status
for each test run.
5. Click
Save As
to save the test file results.
11 - 12
T
EST
TriStar II Plus Operator Manual
303-42800-01 (Rev M ) — Sep 2023