Unit [n] > Krypton Analysis
In 21CFR11 environments, users are required to login to the Confirm application to
start an analysis. Once the analysis window is opened, manual control is disabled
until the analysis has completed. During analysis, pausing and resuming is allowed,
however, steps cannot be skipped.
Krypton analyses are available only if the krypton option is installed.
Data collection is done sequentially — one analysis starts and completes before the
next analysis begins.
Use to perform up to three simultaneous krypton analyses. Use the following guidelines for
performing krypton analyses:
All analyses must use krypton gas. An error message occurs if a gas other than krypton is
n
specified.
From Psat tube must be selected as the Dosing Method on Adsorptive Properties. An error
n
message displays if Normal is selected as the Dosing Method. Samples cannot be
removed from or added to ports until the full set of analyses has completed.
1. Click
Browse
to select a sample file.
2. Click
Report after analysis
plete. On the Report Settings window, select the report destination. Click OK.
3. Click
Start
to start the analysis. A window displays data as they are collected. A short delay
is encountered before the port status changes from the Idle state at the bottom of the
window. When analysis is complete, remove the sample tube and store (or dispose) of the
sample material as applicable.
4. Click
Start
to start the analysis. A window displays data as they are collected. A short delay
is encountered before the port status changes from the Idle state at the bottom of the
window. When analysis is complete, remove the sample tube and store (or dispose) of the
sample material as applicable.
TriStar II Plus Operator Manual
303-42800-01 (Rev M ) — Sep 2023
A
NALYSIS
to automatically generate reports when the analysis is com-

Perform a Krypton Analysis

6 - 13

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TRISTAR II PLUS and is the answer not in the manual?

Table of Contents

Save PDF