Micromeritics TRISTAR II PLUS Operator's Manual page 301

Surface area and porosity analyzer
Hide thumbs Also See for TRISTAR II PLUS:
Table of Contents

Advertisement

    Non-ideality factor:    0.0000620    
  Use isothermal jacket:    No           
         Use filler rod:    No           
       Vacuum seal type:    None         
Degas Conditions
  Degas conditions:    Degas Conditions  
            Smart VacPrep evacuation
     Backfill sample tube:    Automatic         
          Evacuation rate:    0.67 kPa/s        
  Unrest. evacuation from:    0.67 kPa          
             Vacuum level:    1.333224e-03 kPa  
          Evacuation time:    10 min            
    Temperature ramp rate:    10.0 °C/min       
       Target temperature:    30 °C             
            Hold pressure:    13.3 kPa          
                Heating Phase
  Sample                     Ramp             
TriStar II Plus Operator Manual
303-42800-01 (Rev M ) — Sep 2023
F Exported Data Example
F - 2

Hide quick links: