10 Calibration
M
T
ATCH
RANSDUCERS
Unit [n] > Calibration > Match Transducers
This process should not be performed when the main transducer reading is abnormal.
Use to evacuate the system and zero the pressure transducers, then adjust the scale to match
them to the manifold transducer near full scale pressure.
A blank sample tube or small plug must be installed on each selected port prior to start-
ing this process.
1. Ensure that all applicable transducers are selected.
2. Click Start. The window closes when the operation is complete.
3. Click
OK
when the process is complete.
10 - 2
TriStar II Plus Operator Manual
303-42800-01 (Rev M ) — Sep 2023