Perform A Sample Tube Free Space Analysis - Micromeritics TRISTAR II PLUS Operator's Manual

Surface area and porosity analyzer
Hide thumbs Also See for TRISTAR II PLUS:
Table of Contents

Advertisement

R
UN AN ANALYSIS WITH THE UPDATED SAMPLE FILE
1. Go to File > New Sample to create a new sample file or File > Open to open an existing
sample file.
On the Sample Description tab, click the drop-down arrow in the Sample tube field and
select the updated sample tube file to be used for free space analysis.
The Sample mass and Density fields must be correctly entered to calculate the
volume in the sample tube displaced by the sample. This is used by the free
space Calculate option.
2. Go to the Analysis Conditions tab and click
TriStar II Plus Operator Manual
303-42800-01 (Rev M ) — Sep 2023

Perform a Sample Tube Free Space Analysis

Free Space.
6 - 21