Theory Of Operation - Mallinckrodt INOblender 10026 Operation And Maintenance Manual

Inoblender series
Table of Contents

Advertisement

Theory of Operation

1. The INOblender connects to a 3-5 bar (45-75 psig) oxygen supply. The oxygen flow is then
controlled by an integrated oxygen flow meter (2-14 lpm). There is a fixed restrictor in the oxygen
flow path which creates a pressure proportional to the oxygen flow rate.
2. The INOblender is also connected to a 2 bar (30 psig) INOMAX (800 ppm or 400 ppm) supply.
• The INOMAX flow is controlled by a balance regulator in proportion to the oxygen pressure
caused by the fixed restrictor and the INOblender setting.
• The balance regulator maintains a constant INOMAX dose for varying oxygen flow rates.
3. The flows of O
and INOMAX are mixed prior to exiting the INOblender outlet.
2
INOblender Schematic Diagram
INOMAX
Regulator
INOMAX
Cylinder
Balance
Regulator
Dose Knob
INOMAX and
O
Outlet
2
5
Fixed Laminar
Restrictor
INOMAX
Oxygen
Oxygen/INOMAX
Oxygen
Source
Part No. 20732 Rev-02
2014-02

Advertisement

Table of Contents
loading

This manual is also suitable for:

Inoblender 10021Inoblender 10004Inoblender 10070

Table of Contents