SECTION 2: ROUTINE MAINTENANCE

2.1

OVERVIEW

2.2

CLEANING

2.3

PERIODIC SAFETY AND FUNCTIONAL CHECKS

2.4

BATTERY

2.1 Overview
2.2 Cleaning
2.3 Periodic Safety and Functional Checks
2.4 Batteries
The N-20PA requires no routine maintenance, routine service, or calibration. If
service is necessary, contact qualified service personnel or your local
Mallinckrodt representative. Use only Mallinckrodt-approved test equipment
when running a performance test on the N-20PA. The user's institution, local or
national agencies, or both may also require testing.
Dampen a cloth with a commercial, nonabrasive cleaner, and lightly wipe the
surfaces of the N-20PA. Do not spray or pour liquid on the instrument or
accessories. Do not allow liquid to contact connectors, switches, or openings in
the chassis.
The following checks should be performed at least every 2 years by a qualified
service technician.
Inspect the exterior of the N-20PA for damage.
Inspect safety labels for legibility. If the labels are not legible, contact Mallinckrodt
Technical Services Department or your local Mallinckrodt representative.
When the N-20PA is going to be stored for 3 months or more, remove the battery
prior to storage. To replace or remove the battery, refer to Section 5,
Disassembly Guide.
2-1