Tools

Configuration considerations

Steps
2
• Electrostatic protection such as a grounded wrist strap
• T-20 Torx screwdriver
• Diagonal cutter
• 7/16-inch wrench
To achieve a standard factory configuration, it may be
necessary to move an already existing detector. For a
two-detector configuration, factory placement rules are as
follows:
• A thermal conductivity detector (TCD), if present, is always
placed in the front location.
• An FID, if present, is always placed in the back location.
1 Preparing the GC
2 Positioning and Securing the Detector
3 Connecting the Detector
4 Installing the Flow Control Module
5 Installing the Ignitor Extension Cable
6 Installing the Insulation Cup
7 Restoring the GC to Operating Condition
Installation Guide: Flame Ionization Detector (FID)