Installing the Module

Installing the Detector

Installing the Detector

Parts required
Description
Power cord
LAN cable (cross-over or twisted pair network cable)
All modules in the stack should have the latest firmware installed. If other
revisions are required, check with the Agilent support for best match.
Hardware required
Detector (as ordered)
Software required
Appropriate control software or G4208A Instant Pilot (optional).
Preparations
Locate bench space
Provide power connections
Unpack the module
Module is partially energized when switched off, as long as the power cord is
WA R N I N G
plugged in.
Repair work at the module can lead to personal injuries, e.g. shock hazard, when the
cover is opened and the module is connected to power.
Before adding a G1315C/D and G1365C/D into an existing system assure that the existing
N O T E
modules have been updated to firmware revision A.06.02/B.01.02 or above. Otherwise the
ChemStation
For G1315C and G1365C assure that the CompactFlash Card is installed in the rear of the
N O T E
module (required for operation).
50
Make sure that it is always possible to access the power plug.
Remove the power cable from the instrument before opening the cover.
Do not connect the power cable to the Instrument while the covers are removed.
"Performance Specifications"
on page 38 will not recognize modules.