CombiFlash NextGen Systems
Section 2 Preparation
Figure 2-12 Location of the side hole to secure the SLCC ring support

2.9 Position the System

2-12
After completing the various connections, the system can be
moved to its operating position. Turn the system so that the
operator can access all of the front view features and controls
(Figure 1-5). Use care not to damage the connections, tubing, and
cables while moving the system.
Ensure that the CombiFlash NextGen system has at least
3 cm (1.25 inch) of air space behind it for ventilation. Position
the solvent and waste containers as necessary.
Note
When using higher-density solvents such as dichloromethane
(DCM), place the solvent container level with or above the
CombiFlash NextGen unit. Placing solvent containers below
the level of the unit can contribute to decreased flow due to the
high vapor pressure of DCM. This problem becomes more pro-
nounced as the ambient temperature increases.
EAR99 Technology Technology Subject to Restrictions Contained on the Cover Page