Figure 5-8 Line filled with 40 ml of acetone
Figure 5-9 Draining the acetone into a 100 mL
beaker
EAR99 Technology Technology Subject to Restrictions Contained on the Cover Page
6. Place the open end of the P-trap drain line in a 100 mL (or
greater) beaker.
7. Allow the acetone to drain out of the unit by dropping the
end of the P-trap drain line, while it is in the 100 mL bea-
ker, below the P-trap drain line level.
8. Fill the P-trap pump drain line with up to 40 mL of isopro-
pyl alcohol using the same procedure as step 5.
9. Place the open end of the P-trap drain line in a 100 mL (or
greater) beaker.
10. Allow the isopropyl alcohol to drain out of the unit by drop-
ping the end of the P-trap drain line, while it is in the
100 mL beaker, below the P-trap drain line level.
11. Use the D
s menu to set the spray chamber to 60
ETAIL
CombiFlash NextGen Systems
Section 5 Maintenance
.
°C
5-9