CombiFlash NextGen Systems
Section 5 Maintenance
Error/Symptom
No signal on any mass, or very weak
signal
Delayed signal during run.
Sample masses observed but weak, or
masses other than the analyte
observed.
5-16
Table 5-2 General PurIon Troubleshooting
Solution
Verify nitrogen pressure. Nitrogen is required to nebulize the
sample.
Check carrier solvent level, refill carrier solvent and re-prime the
carrier solvent pump. Carrier solvent is needed to bring sample to
the mass spectrometer.
Check that the priming valve is tight. A loose priming valve
prevents sample from getting to the mass spectrometer.
Verify the splitting valve is not leaking. This causes loss of carrier
solvent preventing sample from getting to the mass spectrometer.
Check for leaks around other fittings on the solvent interface.
Verify nitrogen pressure. Sample may drip through the cone and
still be detected later.
Check priming valve tightness. A slight leak will delay sample.
Verify the splitting valve is not leaking.
Check for leaks around other fittings on the solvent interface.
Check polarity to see if the sample ionizes under the opposite
polarity.
Try another ionization settings.
Compound fragments – use fragment mass or try another
ionization setting.
Compound forms adducts.
Compound fails to ionize using the selected probe. Some
compounds ionize better using APCI rather than ESI, for example.