Software Reference - Teledyne CombiFlash NextGen Installation And Operation Manual

Table of Contents

Advertisement

CombiFlash NextGen Systems
Section 3 Operation

3.5 Software Reference

Figure 3-2 NextGen Login screen
3-8
insufficient solvent. Note that isocratic holds and automatic run
extensions during a run will increase the required solvent
volume.
The options available on the alert message will vary depending
on when the condition occurs (before or during a run) and which
method (measured level or estimated usage rate) indicated the
condition.
To clear a solvent level alert message, When solvent level sensing
is enabled on the Instrument Configuration tab in theselect one
of the options below.
• Continue – Select this option after you have added
more solvent to the container. After C
selected, the system measures the solvent level again. If
the level is sufficient to complete the run, the run
continues. If it is not sufficient, the system displays the
alert message again.
• Ignore – This option is available when the estimated
usage rates indicates that the solvent levels are insuffi-
cient. Select I
GNORE
the run. The system then ignores the estimated solvent
usage rate for the remainder of the run, but continues to
measure the solvent level. Should the level become
insufficient, the system will display another alert
message.
• Disable – Select this option to turn off solvent level
measurements and estimated solvent usage rates for the
remainder of the run.
• Cancel – This option appears before the run is started.
Select this option to cancel the run and return to the
main window which will display the current method.
Many features of the software function like a typical PC or tablet
and therefore are not covered in detail. The remainder of this
section covers areas that may be less common or unique to
CombiFlashNextGen systems.
EAR99 Technology Technology Subject to Restrictions Contained on the Cover Page
ONTINUE
to clear the message and continue
is