Power System Problems; Troubleshooting The Autosampler - Teledyne CETAC ASX-112FR Operator's Manual

Autosampler
Table of Contents

Advertisement

7 Troubleshooting the
When the autosampler does not function properly, isolate the problem to
determine if it originates in the host computer, the analytical instrument, the
RS-232 cable, USB cable or the autosampler. If you determine the problem is in
the ASX-112FR, check the power system, the communications interface, or the
sample probe assembly to find the cause of the problem and resolve it.
This chapter explains how to troubleshoot autosampler problems. If you
cannot solve a problem using the steps given in this chapter, you should
contact Teledyne CETAC Technologies Customer Service and Support.

Power System Problems

A possible cause of system malfunction is a problem in the power system. If the
system is not functional, it is possible that it is not receiving power. If this is the
case, the LED status indicator light will be off. To troubleshoot this problem,
complete the following steps in sequence until the problem is solved:
Check that the power switch on the autosampler is turned on.
1
Check that the power cord is plugged in firmly between the power input on the
autosampler and the external power supply and also between the power
2
supply and the wall outlet.
If the cable is plugged in, ensure that it is not damaged in any way.
Check that the LED on the power supply is illuminated.
3
If the LED is not illuminated, check the wall outlet using a device approved for
that purpose.
If the wall outlet is working and the voltage is acceptable, unplug the external
power supply from the autosampler, but leave it plugged into the wall.
 If the LED is not lit, the power supply is faulty and requires replacement.
Autosampler
63