SECTION 2 – OPERATING INSTRUCTIONS

Monitor Cable Connector

This connector is used for monitoring the ultrasonic signals within the test cell. Normally it
is not used. It may be used during system testing.

Top Transducer Connector

The coaxial cable attached to the top transducer must be mated to this connector prior to
beginning a test.

J Thermocouple Connector

The thermocouple that is attached to the top plug must be plugged into this connector prior to
the beginning of a test and any time that it is necessary for the temperature controller to
operate. If the thermocouple is not connected, the temperature controller will display "SBr"
and will not function.

Main Power Connection

Located beside the fuses, it requires a twist lock plug to connect the power. The power
required will be 208-240 VAC, 10A.

Fuses

There are two fuses (2.5A, 250VAC) located in the Instrument Power Receptacle. To check
these fuses, remove the power cord and pry off the access panel with a small screwdriver.

Right Rear Panel

The right rear panel contains all the connections for hydraulic and pneumatic utilities.

Air Inlet, Pump Water Drain/Inlet

These connectors are used to connect the pneumatic and hydraulic utilities.

High Pressure Inlet (optional)

This high-pressure connector exists so that pressure may be controlled with an
alternate pressure control system.