1.8 Preventative Maintenance and Cleaning

Make sure that the instrument is in a clean, dry working environment (dust-free, moderate temperature,
low humidity, etc.).
Make sure the instrument is on a level surface.
Hands/fingers must be clean and free of sample residue. Not doing so may result in deposit build-up on
the upper part of the shaft and cause interference between the shaft and the pivot cup.
Be sure to remove the spindle from the instrument prior to cleaning. Note left-handed thread. Severe
instrument damage may result if the spindle is cleaned in place.
When cleaning, do not apply excessive force, which may result in bending spindles.
Instrument and Display:
Immersed Components (spindles):
M21-400 REV.1.1
The Status Bar provides information relating to the date and time (as configured by the user)
and various connections to the DVPlus Viscometer.
The Title Bar identifies the activity to be conducted in the current view and includes any
navigation icons that are relevant.
The Test Parameters are where the user enters the test factors relevant for the test to be
conducted.
The Static Readings are the DVPlus data while at rest.
The Command Keys indicate the action that can be taken. These keys will vary from view to
view depending on what actions are relevant.
D V P L U S D I G I TA L V I S C O M E T E R - O P E R AT I O N A L M A N U A L
Clean with a dry, non-abrasive cloth. Do not use solvents
or cleaners. The instrument housing is manufactured from
polycarbonate ABS. Clean Instrument housing with mild soap
and water. Do not apply solvent to the instrument!
Spindles are made of stainless steel. Clean with a
non-abrasive cloth and solvent appropriate for the sample
material.