Slope Trigger - UNI-T MSO/UPO2000 Series Programming Manual

Digital phosphor oscilloscope
Table of Contents

Advertisement

Programming Manual
This command is used to set video standard.
Return format:
Query returns { NTSC | PAL | SECAM }.
For example:
:TRIGger:VIDeo:STANdard NTSC
:TRIGger:VIDeo:STANdard?
:TRIGger:VIDEO:LINE
Command format:
:TRIGger:VIDEO:LINE <value>
:TRIGger:VIDEO:LINE?
Functional description:
This command is used to set the assigned line. <value> represents the assigned line, range is relative to
video standard.
Return format:
Query returns the current assigned line.
For example:
:TRIG:VIDEO:LINE 50
:TRIG:VIDEO:LINE?

Slope Trigger

:TRIGger:SLOPe:QUALifier
Command format:
:TRIGger:SLOPe:QUALifier {GREaterthan | LESSthan | INRange}
:TRIGger:SLOPe:QUALifier?
Functional description:
This command is used to set the setting condition of slop time, which includes GREaterthan (greater than),
LESSthan (less than) and INRange (between).
Return format:
Query returns {GREaterthan | LESSthan | INRange}.
For example:
:TRIGger:SLOPe:QUALifier GRE
:TRIGger:SLOPe:QUALifier?
:TRIGger:SLOPe:SLOPe
Command format:
:TRIGger:SLOPe:SLOPe {POSitive|NEGative}
:TRIGger:SLOPe:SLOPe?
Functional description:
Instruments.uni-trend.com
Set video standard to NTSC.
Query returns NTSC.
Set the assigned line of video synchronization to 50.
Query returns 50.
Set slop condition to GREaterthan.
Query returns GREaterthan.
MSO/UPO2000 Series
45 / 156

Advertisement

Table of Contents
loading