[Test] Measurement Page

4.1
<Test>page
<Test> page is mainly used to display the measurement and sorting results. Three
common functions can be set on this page, including:
TRIGGER - The Trigger Mode of the Measurement
FUNC – Parameters of Measurement
RANGE – Setting and Display of Resistance and Voltage Range

【Trigger】

4.1.1
The instrument has 2 trigger modes: internal trigger and external trigger (including manual
/remote)

Function

INT
EXT
Setting Steps
1.
Press【Test】key to enter the main page of <Test>;
Use the cursor key▼to move the cursor to the 【TRIGGER】 field;
2.
According to user's needs, use the function keys at the bottom of the screen to
3.
select the desired trigger mode.
【Function】
4.1.2
The instrument has 3 measurement functions: RV, R and V.
Function
RV
R
V
Setting Steps
1.
Press【Test】key to enter the main page of <Test>;
Use the cursor key▼to move the cursor to the 【FUNC】field;
2.
According to user's needs, use the function keys at the bottom of the screen to select the
3.
desired measurement items.
Figure 4-1<Test> Page
Also called continuous test, the trigger signal is continuously tested by the
internal of the instrument according to inherent cycle. Generally choose this
method for measurement
Maunal:Each time press the【Trigger】key,the instrument will execute a
measurement cycle, and the instrument is in a waiting state at other times.
Remote:Send the command TRIGGER to measure once and return the
measured value.
Measure and display resistance and voltage values at the same time
Only measure and display the resistance value
Only measure and display the voltage value
Description
Description
17
UT3550 Battery Tester Users Manual