UNI-T MSO/UPO2000 Series Programming Manual page 98

Digital phosphor oscilloscope
Table of Contents

Advertisement

Programming Manual
:SBUS:CANFD:FRAMe:TYPE
Command format:
:CANFD
:SBUS
:CANFD
:SBUS
Functional description:
This command is used to set CAN-FD bus trigger frame type.
Return format:
Query returns { DATA | REMote | OVERload | ERRor }.
For example:
:CANFD
:SBUS
Set CAN-FD bus trigger frame type to ERRor.
:CANFD
:SBUS
:SBUS:CANFD:ID:MODE
Command format:
CANFD
:SBUS:
CANFD
:SBUS:
Functional description:
This command is used to set ID identifier frame format of CAN-FD bus of the oscilloscope.
Return format:
Query returns {STANdard | EXTended | FD_STD| FD_EXT }.
For example:
CANFD
:SBUS:
Set ID identifier frame format of CAN-FD bus to STANdard.
CANFD
:SBUS:
:SBUS:CANFD:ID
Command format:
CANFD
:SBUS:
CANFD
:SBUS:
Functional description:
This command is used to set ID identifier frame data of CAN-FD bus of the oscilloscope. Set the
corresponding ID identifier according to the mode of :SBUS:CAN:ID:MODE. Binary character string data
presented by parameter is 0, 1 or X. X represents unsure.
Standard frame range is 0x0~0x7FF, data bit occupies 11bits; Extend frame range is 0x0~0x1FFFFFFF, data
bit occupies 29bits; FD standard frame range is 0x0~0x7FF, data bit occupies 11bits; FD extend frame range
is 0x0~0x1FFFFFFF, data bit occupies 29bits.
Return format:
Query returns binary character data.
Instruments.uni-trend.com
:FRAMe:TYPE { DATA | REMote | OVERload | ERRor }
:FRAMe:TYPE?
:FRAMe:TYPE ERRor
:FRAMe:TYPE?
:ID:MODE {STANdard | EXTended | FD_STD| FD_EXT }
:ID:MODE?
:ID:MODE STANdard
:ID:MODE?
:ID <string>
:ID?
Query returns ERRor.
Query returns STANdard.
MSO/UPO2000 Series