UNI-T MSO/UPO2000 Series Programming Manual page 25

Digital phosphor oscilloscope
Table of Contents

Advertisement

Programming Manual
:FUNCtion:FFT:
Command format:
:FUNCtion:FFT:DETEction
:
FUNCtion:FFT:DETEction:AVERage
Functional description:
This command is used to set the average number of times of average spectrum. The range is 2^n, n take
value from 1~10.
Return format:
Query the average number of times of average spectrum.
For example:
:FUNCtion:FFT:DETEction
Set the average number of times of average spectrum to 64.
:FUNCtion:FFT:DETEction
:
FUNCtion:FFT:DETEction:MAXHold
Command format:
:FUNCtion:FFT:DETEction:MAXHold {OFF|PPEAK|NPEAK| AVERage|SAMPle}
FUNCtion:FFT:DETEction:MAXHold ?
:
Functional description:
This command is used to set the detection mode of the maximum hold spectrum.
OFF: Turn off the average spectrum;
PPEAK:Take the maximum value within the range of each sampling point;
NPEAK:Take the minimum value within the range of each sampling point;
AVERage:Take the average value within the range of each sampling point;
SAMPle:Take the value of first point within the range of each sampling point.
Return format:
Query returns
For example:
:FUNCtion:FFT:DETEction:MAXHold PPEAK
the detection mode of the maximum hold spectrum to +peak detection.
Set
:FUNCtion:FFT:DETEction:MAXHold?
Query returns PPEAK.
:
FUNCtion:FFT:DETEction:MINHold
Command format:
:FUNCtion:FFT:DETEction:MINHold {OFF|PPEAK|NPEAK| AVERage|SAMPle}
FUNCtion:FFT:DETEction:MINHold ?
:
Functional description:
This command is used to set the detection mode of the minimum hold spectrum.
OFF: Turn off the average spectrum.
PPEAK:Take the minimum value within the range of each sampling point;
NPEAK:Take the minimum value within the range of each sampling point;
Instruments.uni-trend.com
:AVERage:COUNt
DETEction
:AVERage:COUNt
:AVERage:COUNt 64
:AVERage:COUNt
the detection mode of the maximum hold spectrum.
<value>
:COUNt
?
?
Query returns 64.
MSO/UPO2000 Series
25 / 156