Query returns 2.000e003, unit is Hz.
:MEASure:RISetime?
Command format:
:MEASure:RISetime? [<source>]
Functional description:
This command is used to measure rising time of the specified channel waveform. <source> take value from
CHANnel1, CHANnel2, CHANnel3 or CHANnel4. Omitting represents the current channel.
Return format:
Query returns 5.000e-005, unit is s.
:MEASure:FALLtime?
Command format:
:MEASure:FALLtime? [<source>]
Functional description:
This command is used to measure falling time of the specified channel waveform. <source> take value from
CHANnel1, CHANnel2, CHANnel3 or CHANnel4. Omitting represents the current channel.
Return format:
Query returns 5.000e-005, unit is s.
:MEASure:PERiod?
Command format:
:MEASure:PERiod? [<source>]
Functional description:
This command is used to measure period of the specified channel waveform. <source> take value from
CHANnel1, CHANnel2, CHANnel3 or CHANnel4. Omitting represents the current channel.
Return format:
Query returns 5.000e-003, unit is s.
:MEASure:PWIDth?
Command format:
:MEASure:PWIDth? [<source>]
Functional description:
This command is used to measure positive pulse width of the specified channel waveform. <source> take
value from CHANnel1, CHANnel2, CHANnel3 or CHANnel4. Omitting represents the current channel.
Return format:
Query returns 5.000e-003, unit is s.
:MEASure:NWIDth?
Command format:
:MEASure:NWIDth? [<source>]