Trigger Command; Trigger Control - UNI-T MSO/UPO2000 Series Programming Manual

Digital phosphor oscilloscope
Table of Contents

Advertisement

Programming Manual
edge of <source2>. <source> take value from CHANnel1, CHANnel2, CHANnel3 or CHANnel4.
Return format:
Query returns 5.000e-003, unit is s.

TRIGger Command

This command is used to control trigger sweep mode and trigger specification. Trigger determines when the
oscilloscope to start sampling data and display waveform.

Trigger Control

:TRIGger:MODE
Command format:
:TRIGger:MODE <mode>
:TRIGger:MODE?
Functional description:
This command is used to set trigger mode and it can automatically suitable for trigger mode of different
model.
<mode> contains EDGE (edge trigger), PULSe (pulse width trigger), VIDeo (video trigger), SLOPe (slope
trigger), RUNT (runt trigger), WINDow (over-amplitude trigger), DELay (delay trigger), TIMeout (timeout
trigger) DURation (duration time trigger),SHOLd (setup hold trigger), NE (Nth edge trigger) and PATTern
(pattern trigger).
Return format:
Query returns the trigger mode.
For example:
:TRIGger:MODE NE
:TRIGger:MODE?
:TRIGger:FORCe
Command format:
:TRIGger:FORCe
Functional description:
This command is sutiable for the oscilloscope if there is no properly trigger condition. Executing this
command to force to produce a trigger signal to generate input waveform and display it.
For example:
:TRIG:FORC
:TRIGger:SWEep
Command format:
Instruments.uni-trend.com
Set trigger mode to Nth edge trigger.
Query returns NE.
Force trigger
MSO/UPO2000 Series
39 / 156

Advertisement

Table of Contents
loading

Table of Contents