UNI-T UTD2052CL+ User Manual

UNI-T UTD2052CL+ User Manual

Digital oscilloscope
Hide thumbs Also See for UTD2052CL+:
Table of Contents

Advertisement

UTD2000 Series
Digital Oscilloscope
User Manual
www.uni-trend.com.cn

Advertisement

Table of Contents
loading

Summary of Contents for UNI-T UTD2052CL+

  • Page 1 UTD2000 Series Digital Oscilloscope User Manual www.uni-trend.com.cn...
  • Page 2: Guaranty And Declaration

    Preface Dear custumer, Thank you for choosing this UNI-T device. In order to safely and correctly use this instrument .please read this manual thoroughly, especially the Safety Notes part. After reading this manual, it is recommended to keep the manual at an easily accessible place, preferably close to the device, for future reference.
  • Page 3 The probe, other accessories and fuses are not covered by warranty. If any genuine defect is found during the valid warranty period, Uni-T has the option to repair the defective product without any charge for parts or labor, or replace it with another product (at the discretion of Uni-T).
  • Page 4: General Safety Overview

     Avoid fire and personal injury:  Use the correct power line: Only use the dedicated UNI-T power supply appointed to the local region or country for this product. The rating shall not be lower than the rating requirements. Mains supply cord set specification...
  • Page 5 UTD2000 Series User Manual is connected.  Do not operate the product if you suspect it is faulty, and please contact UNI-T authorized service personnel for inspection. Any maintenance, adjustment, or replacement of parts must be performed by UNI-T authorized maintenance personnel.
  • Page 6: Safety Terms And Symbols

    UTD2000 Series User Manual Safety Terms and Symbols The following terms may appear in this manual: Warning: The conditions and behaviors may endanger life. Note: The conditions and behaviors may cause damage to the product and other properties. The following terms may appear on the product: Danger: Performing this operation may cause immediate damage to the operator.
  • Page 7: Preface

    UTD2000 Series User Manual Preface The Manual introduces information related to operation of digital storage oscilloscope of UTD2000 series user manual. The Manual comprises the following chapters: Chapter 1 Introduction Guide Chapter 2 Vertical channel setting Chapter 3 Horizontal system setting Chapter 4 Trigger system setting Chapter 5 Sampling system setting Chapter 6 Display system setting...
  • Page 8 UTD2000 Series User Manual powerful functions required for faster completion of measurement tasks. User can observe faster signals on UTD200 series DSO by high-speed real time sampling and equivalent sampling. Clear LCD and mathematical operation functions make it easier for users to observe and analyze signal problems in a faster and clearer way.
  • Page 9: Table Of Contents

    UTD2000 Series User Manual Contents Guaranty and Declaration ....................2 General Safety Overview ..................... 4 Safety Terms and Symbols....................6 Preface..........................7 Contents ..........................9 Chapter I Introductions ....................11 1.1 General Inspection ....................11 1.2 Function Inspection ....................11 1.3 Front and Rear Panel ...................
  • Page 10 UTD2000 Series User Manual 7.2 Voltage Parameter ....................38 7.3 Time parameter ....................39 Chapter VIII Cursor Measurement .................. 41 Chapter IX Storage and Load System ................41 9.1 Setup storage and load ..................43 9.2 Waveform storage and load ................. 44 9.3 Bit map storage and load ..................
  • Page 11: Chapter I Introductions

    In case of damages to the appearance of instrument, abnormal operation or failure to pass the performance tests, please contact the UNI-T product distributors or local UNI-T offices. If the instrument is damaged due to transportation, please pay attention to keeping the package, notify transportation department and UNI-T product distributors and UNI-T will make arrangement for repair or replacement.
  • Page 12 UTD2000 Series User Manual frequency defaults at 1kHz. Probe Compensation Siganl Tip Ground Terminal (4) Function Check Press AUTO (auto setting) button, square wave appears on the display screen, the wave range is about 3Vpp and the frequency is 1kHz. Go back to step 3 and check other channels in a same way.
  • Page 13: Front And Rear Panel

    UTD2000 Series User Manual Warning: In order to avoid electric shock when measuring high voltage with probe, please ensure that the insulation lead of the probe is in good condition and do not contact the metal part of the probe when connecting high pressure power supply. 1.3 Front and Rear Panel Control Menu button Function menu button...
  • Page 14: Automatic Settings Of Waveform Display

    UTD2000 Series User Manual 1.4 Display (1) Display description Horizontal trigger area Main time base setting Time of center scale line Trigger state display Soft button menu corresponds to different buttons Channel 1 mark Waveform display window Channel 2 mark Vertical scale coefficient of channel (2) Soft button Press any soft button to activate corresponding menu.
  • Page 15: Introduction To The Horizontal System

    UTD2000 Series User Manual control area. Practices below will gradually guide you to get familiar with the controlling of the vertical system. Figure 1-6 Vertical Control Area on the Panel (1) Vertical POSITION: vertical position knob, user can change the current vertical position of channel waveform, the vertical position value will display on the baseline cursor area.
  • Page 16: Introduction To The Trigger System

    UTD2000 Series User Manual 1. Use horizontal SCALE knob to change gear position settings of horizontal time base and observe the changes of state information. Turn horizontal SCALE knob to change “SEC/DIV” time base gear position, and you can find that corresponding changes have taken place in the time base gear position display of corresponding channel in the status bar.
  • Page 17: Chapter Ii Vertical System

    UTD2000 Series User Manual Chapter II Vertical System UTD2000 series DSO provides two analog input channels, each channel has an independent vertical menu. Each channel for UTD2000 series DSO can be set independently through the vertical system menu. After pressing CH1 or CH2 function key, the system shall display function menu of CH1 or CH2 channel.
  • Page 18: Setting Channel Bandwidth Limited

    UTD2000 Series User Manual Figure 2-1 Blocked DC Component of Signal Press F1 to select as DC coupling, you should be able to see both the DC and AC components of measured signals at CH1 channel as shown in figure below. Figure 2-2 Simultaneous Display of Signal DC and AC Components If you Press F1 to select as GND, to set CH1 to connect to the internal ground of the instrument, both DC and AC components of the input signal are blocked.
  • Page 19: Setting Probe Rate

    UTD2000 Series User Manual Figure 2-4 Waveform Display When Setting Bandwidth Limit as OFF Press F2 to set the BW Limit as 20MHz, the noises or high-frequency component over 20MHz in measured signal shall be attenuated, waveform display shown in figure below. BW Limit Mark Figure 2-5 Waveform Display When Setting Bandwidth throttling as ON 2.3 Setting probe Rate...
  • Page 20: Settings Of Waveform Inverting

    UTD2000 Series User Manual Adjustments of vertical deflection factor V/div gear position can be set to coarse adjustment and fine adjustment. For coarse adjustment, V/div can be adjusted in step of multiple of 1, 2 and 5, the V/div range is 1mV/div~20V/div. Fine adjustments refer to changing deflection factor with smaller stepping within the current vertical gear position range.
  • Page 21: Mathematical Operation Function

    UTD2000 Series User Manual 2.6 Unit Press UNIT to set the unit as “V” or “A”, the default unit is V. After the unit is set, the unit on channel status tag will change accordingly. 2.7 Mathematical operation function Display results of CH1 & CH2 waveform mathematic operation (+、–、×、÷ ). Figure 2-10 Math Operation Table 2-3:Math Menu Function Menu...
  • Page 22 UTD2000 Series User Manual CH1 or CH2 Hanning Set to select Hanning, Hamming, Hamming Window Blackman, or Rectangle window Blackman function Rectangle Vrms Set vertical units as Vrms or Vertical unit dBVrms dBVrms Select FFT window Assuming that YT waveform is continuously repeated, DSO shall conduct FFT conversion for time record with finite length.
  • Page 23 UTD2000 Series User Manual 2.9 Digital Filter Filter the frequency of specified waveband in signals by adjusting upper/lower frequency limit. Table 2-6 Digital Fiter Menu Function menu Setting Description Type Digital filter Digital filter Set CH1 as operation waveform Source Set CH2 as operation waveform Only when the signal frequency is lower than current upper frequency...
  • Page 24: Chapter Iii Horizontal System

    UTD2000 Series User Manual Chapter III Horizontal System 3.1 Horizontal control (1) Horizontal control knob Change the horizontal time-base scale through SCALE knob, changing of horizontal scale can lead to expansion or shrink of waveforms relative to screen center. The horizontal system can be controlled by the following panel buttons/knobs.
  • Page 25: Window Extension

    UTD2000 Series User Manual 3.3 Window extension Extension window is used to enlarge the waveform so that users view the image details. The setting of window extension shall not be slower than that of main time base. Waveform horizontal extension part Extensive time base Main time base Waveform horizontal extension part...
  • Page 26 UTD2000 Series User Manual According to sin θ=A/B or C/D, θ is the angle difference of channels, the definition of A, B, C, D is shown in figure above, so the angle difference is: θ=±arcsin (A/B) or θ=±arcsin (C/D). If the elliptical axis lies on I, III quadrant, then the phase difference of angle should lie on I, IV quadrant, that is within (0~π/2) or (3π/2~2π).
  • Page 27: Chapter Iv Trigger System

    UTD2000 Series User Manual Chapter IV Trigger System When to collect data and display waveform depends on the trigger system. The DSO will convert the unstable displays into meaningful waveform after the trigger is set correctly. When the DSO start collecting data, it will firstly collect enough data to draw a waveform on the left of trigger.
  • Page 28: Pulse Width Trigger

    UTD2000 Series User Manual Set to automatic trigger. The DSO will continuously perform data acquisition AUTO without triggering signal. Set to normal trigger. The DSO will only perform data acquisition when there is Mode Normal triggering signal. Set to single trigger. The DSO will only perform 1 cycle of the data acquisition Single when there is triggering signal.
  • Page 29 UTD2000 Series User Manual Table 4-3 Pulse width setting menu Function Menu Setting Description Positive Set the positive pulse width as trigger signal Polarity Negative Set the negative pulse width as trigger signal > Trigger when pulse width is greater than the set value Pulse width <...
  • Page 30: Video Trigger

    UTD2000 Series User Manual The DSO generates a trigger when the slew rate of rising edge of signal is Rising < lower than the specified slew rate. The DSO generates a trigger when the slew rate of rising edge of signal is Rising <>...
  • Page 31: Alternating Trigger

    UTD2000 Series User Manual NTSC Applicable NTSC video signal. All line Set the video line to synchronous triggering. Set the video to sync and trigger on specific line, adjust through the Line multifunctional knob on front panel. Sync Odd field Set the video to sync and trigger on odd field.
  • Page 32 UTD2000 Series User Manual Figure 4-3 Observing Signals with Two Different Frequencies by Alternating Trigger Mode Table 4-7 Alternating Trigger Menu (Page 1) Function Setting Description Menu Type Alternating Select CH1 as current channel Channel selection Select CH2 as current channel Edge Set the edge as trigger mode Trigger mode...
  • Page 33: Explanation Of Nouns

    UTD2000 Series User Manual 4.6 Trigger Holdoff Trigger holdoff time adjustment is used to observe complex waveform (pulse string series). Holdoff time refers to the waiting time for the DSO to restart trigger circuit. During holdoff time, the DSO will not trigger until the holdoff time is over. Table 4-9 Trigger Holdoff Menu Function Menu Setting...
  • Page 34 UTD2000 Series User Manual ■ External trigger: This type of trigger source can be used not only in collecting data in two channels but also in triggering the third channel at the same time. For example, the external clock or the signal from circuit to be measured can be served as trigger source. EXT trigger source use the external trigger signal connected to EXT TRIG junction.
  • Page 35: Chapter V Acquire System

    UTD2000 Series User Manual Chapter V Acquire System As shown in the figure below, the ACQUIRE on the control area is the function button of Acquire System. Figure 5-1 Function Key of Acquire System Press ACQUIRE button to enter data acquisition setting menu, set the data acquisition mode through menu control button.
  • Page 36 UTD2000 Series User Manual Figure 5-1 Non-average acquisition method Figure 5-2 Average acquisition of 8 numbers Notice: (1) Please select real-time sampling mode when observing single signal. (2) Please select equivalent sampling mode when observing high frequency periodic signal. (3) Please select peak detection mode when hoping to observe the signal envelope for any modulation signal.
  • Page 37: Chapter Vi Display System

    UTD2000 Series User Manual Chapter VI Display System As shown in the figure below, the DISPLAY on the control area is the function button of Display System. Figure 6-1 Function Keys of Display System Press DISPLAY button to enter display setting menu as shown in the table below. Table 6-1 Display Menu (Page 1) Function Setting...
  • Page 38: Chapter Vii Measurement Parameters

    UTD2000 Series User Manual Chapter VII Automatic Measurement UTD2000 series DSO support up to 34 types of waveform measurement parameters. Figure 7-1 Function Key of Auto Measurement 7.1 Measurement Menu Press MEASURE to enter Measurement menu. Table 7-1 Automatic measurement menu (Page 1) Function Menu Setting Description...
  • Page 39: Voltage Parameter

    UTD2000 Series User Manual 7.2 Voltage Parameter Figure 7-2 Voltage Parameter Diagram Voltage parameters of UTD2000 series DSO include: Max value (Max): The voltage value from waveform highest level to GND. Min value (Min): The voltage value from waveform lowest level to GND. Top value (High): The voltage value from waveform flat top to GND.
  • Page 40 UTD2000 Series User Manual Frequency (Freq): The reciprocal of the period Rise time (Rise): The time for waveform amplitude to rise from 10% to 90%. Fall time (Fall): The time for waveform amplitude to fall from 90% to 10%. Rise delay: The time for main source and slave source to delay between two rising edges Fall delay: The time for main source and slave source to delay between two falling edges.
  • Page 41: Chapter Viii Cursor Measurement

    UTD2000 Series User Manual Chapter VIII Cursor Measurement Use CURSOR to measure waveform X axis (time) and Y axis (voltage). Press CURSOR to enter cursor measurement menu. Figure 8-1 Function Key of CURSOR 8.1 Cursor measurement menu Press CURSOR to enter cursor measurement menu. Table 8-1 cursor measurement menu Function menu Setting...
  • Page 42 UTD2000 Series User Manual Figure 8-2 Cursor measurement for time X time: Rotate Multipurpose knob to adjust the position of AX, press Multipurpose knob to adjust BX, shit between BX and AX with same way. A/B represents the time distance between zero point and the cursor A/B. B-A represents the difference beween cursor A and cursor B.
  • Page 43: Setup Storage And Load

    UTD2000 Series User Manual Chapter IX Storage The DSO setting, waveform and screen image can be store in DSO or USB, the stored setting or waveform can be recalled when needed. Press STORAGE to enter storage setting interface. Figure 9-1 Function Key of STORAGE 9.1 Set storage and load Press the Storage button, then Press F1 to choose storage type Setup, and you will see the following menu.
  • Page 44: Waveform Storage And Load

    UTD2000 Series User Manual 9.2 Waveform storage and load Press the Storage button, then press F1 to choose storege type as Reference Waveform, and you will see the following menu. Table 9-2 Reference waveform menu (Page 1) Function Menu Setting Description Reference Type...
  • Page 45: Bit Map Storage And Load

    UTD2000 Series User Manual Storage Press the button to store. 9.3 Bit map storage and load Press the Storage button, then press F1 to choose storege type Data File, and you will see the following menu. Table 9-6 Interface storage menu Function Menu Setting Description...
  • Page 46: Chapter X Utility System

    UTD2000 Series User Manual Chapter X Utility System Press UTILITY to enter utility menu. Figure 10-1 Function Key of UTILITY Table 10-1 Utility menu (Page 1) Function menu Setting Description System Config See chart 10-3 Enter system setup menu, the operation option include self-clibration, system information and clear information.
  • Page 47: Pass/Fail

    UTD2000 Series User Manual 10.1 Pass/Fail Pass/Fail test: Detect whether or not the input signal is in the specified scope of template. If the input signal is within the scope, it is PASS, while if the input signal is out of the scope, it is FAIL.
  • Page 48: Recorder

    UTD2000 Series User Manual tolerance. Load Load the reference waveform. Horizontal 1~255 Set horizontal tolerance of template through multifunctional knob. tolerance Vertical 1~255 Set vertical tolerance of template through multifunctional knob. tolerance Back Return to previous menu (Pass/Fail Test Menu). (8) Start the test, press F1 to conduct Pass/Fail test.
  • Page 49 UTD2000 Series User Manual 10.3 AUTO strategy As mentioned before, press AUTO button to activate the function of Waveform Auto Setting. To get a best diplay waveform, DSO automatically adjust vertical scale, horizontal time base and trigger mode based on input signals. This DSO allow users to set relevant parameters of auto setting function.
  • Page 50: Chapter Xi Other Funtion Botton

    UTD2000 Series User Manual Chapter XI Other function botton 11.1 AUTO Setting Based on input signals, Auto Setting selects appropriate time-based scale, Volt/Divs, trigger based to display waveform on the screen automatically. Press AUTO to enter automatic setting. AUTO Setting is only applicable to conditions below: 1) AUTO Setting is only applicable to signals with simple and single frequency.
  • Page 51: Help Menu

    UTD2000 Series User Manual 11.3 Help Menu Press the HELP key to go into help menu, then press any key will display the help information of this key. 11.4 Upgrade program The USB upgrade program makes upgrade easier and more flexible. To use this function, please follow the steps below: (1) Download a program file to be updated from internet and store the file in USB.
  • Page 52: Chapter Xii Applicaion Example

    UTD2000 Series User Manual Chapter XII Application example Example 1: Measuring simple signals To observe and measure an unknown circuit signal, and to quickly display and measure the signal's frequency and peak-to-peak value. (1) To quickly display this signal, follow the steps below: ①...
  • Page 53: Example 3: Acquiring Single Signals

    UTD2000 Series User Manual (1) To display CH1 and CH2 signals ① Press AUTO. ② Continue to adjust the horizontal and vertical range until you get the desired waveform display. ③ Press CH1 to select CH1. Adjust vertical position of the CH1 waveform by turning the vertical position control knob.
  • Page 54: Example 4: Reducing Random Noise Of Signals

    UTD2000 Series User Manual ① Press TRIG MENU in the trigger control zone to display the trigger setup menu. ② In this menu, use F1~F5 set the trigger type to EDGE, set trigger source to CH1, set inclination to Rising, set trigger type to Single and set trigger coupling to AC. ③...
  • Page 55: Example 5: Using The Cursors For Measurement

    UTD2000 Series User Manual (2) Connect the signal to ensure stable display of waveform. (3) Improving trigger by setting trigger coupling. ① Press TRIG MENU in the trigger zone to display the trigger setup menu. ② Set trigger coupling to Low Frequency Holdoff or High Frequency Holdoff. Low frequency holdoff is seting up a high-pass filter.
  • Page 56 UTD2000 Series User Manual Measuring one step voltage of the step signal To measure one step voltage of the step signal, follow the steps below: 1. Press CURSOR to display the cursor measurement menu. 2. Press menu operation key F1 to set cursor type to VOLTAGE. 3.
  • Page 57: Chapter Xiii System Prompts And Trouble-Shooting

    UTD2000 Series User Manual Chapter XIII System Prompts and Trouble- shooting 13.1 Definitions of System Prompts Adjustment at Ultimate Limit: This informs you that the multifunction control knob has reached its adjustment limit in the current status. No further adjustment is possible. When the vertical deflection factor switch, time base switch, X shift, vertical shift and trigger level adjustments have reached their ultimate limits, this prompt will appear.
  • Page 58 UTD2000 Series User Manual or to filter any high or low frequency noise that is interfering with triggering. (4) Slow refresh ①. Check if the acquiring mode in ACQUIRE button menu is average, and the average times are large. ②. If increasing the refresh speed is needed, please decrease the average times appropriately or select other acquiring modes, i.e.
  • Page 59: Chapter Xiv Technical Information

    UTD2000 Series User Manual Chapter XIV Technical Information Except for those specifications marked with “Typical”, all specifications have warranties. Unless otherwise specified, all technical specifications are applicable to the probes with attenuation switch set as 10× as well as UTD2000 series DSO. DSO must first meet the following two conditions to meet those specification standards: ...
  • Page 60 UTD2000 Series User Manual Measurement Single time: ± (1 sampling time interval+50ppm×reading+0.6ns) accuracy of time >16 average values: ± (sampling time interval+50ppm×reading+0.4ns) interval (△T) (full bandwidth) Vertical UTD2202 UTD2052 UTD2072 UTD2102PRO UTD2102 UTD2152 CEX+ Model UTD2152CL UTD2102CL+ CEX+ UTD2202 Analog Bandwidth 50MHz 70MHz 100MHz...
  • Page 61 UTD2000 Series User Manual quantity of ≥16, the voltage difference (ΔV) between any two points on the waveform: ± accuracy of voltage difference (3%×reading+0.05div) (△V) (average sampling mode) Trigger System Specifications ≤1div Trigger sensitivity Interior:From the screen center ± 10div Range of trigger level EXT:...
  • Page 62 UTD2000 Series User Manual Measurements Manual mode Voltage difference between cursors (△V), Time difference between cursors (△T), Reciprocal of △T (Hz) (1/△T)) Cursor Track mode Voltage value and time value of point of waveform. Auto measurement Cursor display is allowed on auto measurement mode. mode Vpp, Vamp, Vmax, Vmin, Vtop, Vbase, Vmid, Average, Vrms, Overshoot, Preshoot, Frequency, Period, Automatic...
  • Page 63 UTD2000 Series User Manual Display Displays types LCD with Diagonal of 178mm (7-inch) Display resolution 800 horizontal×RGB× 480 vertical pixels Display color Color Waveform luminance Adjustable Backlight intensity (Typical) 300nit Language Chinese and English Interface function: Standard:USB-Host, USB-Device, EXT Trig, Pass/Failt. Standard configuration Option:Multimeter module (UT-M12), LAN.
  • Page 64: Chapter Xv Appendix

    Warnings: Please make sure that the instrument is completely dry before powering on again, to avoid electrical short circuit or injury. Appendix C Warranty UNI-T (Uni-Trend Technology (China) Co., Ltd.) warrants that product produced and marketed by it will be free from defects in materials and workmanship for a period of three...
  • Page 65: Appendix D Contact Us

    If any such product proves defective during this warranty period, UNI-T will repair the defective product or provide a replacement according to specific terms and conditions of the warranty. To request maintenance and repair service or a full copy of the warranty, please contact your nearest UNI-T sales and maintenance office.