UNI-T MSO/UPO2000 Series Programming Manual page 114

Digital phosphor oscilloscope
Table of Contents

Advertisement

Programming Manual
This command is used to set output amplitude of the specified channel.
<amp> represents voltage, unit is assigned by the current channel. (1mVpp ~ maximum value of output
under the current load)
If the current unit is VPP, then the maximum value under the current load = the current load *20/(50+ the
current load)
<n>: channel number, n take value from 1, 2.
Return format:
Query returns output amplitude of the specified channel in scientific notation.
For example:
AWG:CHANnel1:BASE:AMPLitude 2
AWG:CHANnel1:BASE:AMPLitude?
AWG:CHANnel<n>:BASE:PERiod
Command format:
AWG:CHANnel<n>:BASE:PERiod <period>
AWG:CHANnel<n>:BASE:PERiod?
Functional description:
This command is used to set output period of the specified channel.
<period> represents period, unit is s.
If it is sine wave, the range is the current allowed maximum time ~ 1e3s.
<n>: channel number, n take value from 1, 2.
Return format:
Query returns the amplitude upper limit of the specified channel in scientific notation.
For example:
AWG:CHANnel1:BASE:PERiod 0.002
AWG:CHANnel1:BASE:PERiod?
AWG:CHANnel<n>:BASE:PHASe
Command format:
AWG:CHANnel<n>:BASE:PHASe <phase>
AWG:CHANnel<n>:BASE:PHAse?
Functional description:
This command is used to set output phase of the specified channel.
<phase> represents phase, unit is°. The range is -360~360.
<n>: channel number, n take value from 1, 2.
Return format:
Query returns output phase of the specified channel.
For example:
AWG:CHANnel1:BASE:PHAse 20
AWG:CHANnel1:BASE:PHAse?
Instruments.uni-trend.com
Set output amplitude of channel 1 to 2 V.
Query returns 2e+0.
Set output period of channel 1 to 2ms.
Query returns 2e-3.
Set output phase of channel 1 to 20°.
Query returns 20.
MSO/UPO2000 Series
114 / 156