Trigger Holdoff - UNI-T UPO2102 User Manual

Digital phosphor oscilloscope
Hide thumbs Also See for UPO2102:
Table of Contents

Advertisement

User's Manual
Note:When Multi-Scopes is split screen, horizontal of each screen is still 700 pixels, if the there are four
split screens, horizontal of each screen only has 350 pixels.

6.6 Trigger Holdoff

Trigger holdoff is used for viewing complicated waveform (such as pulse string). Holdoff time refers to the
time that the oscilloscope waits to restart the trigger circuit. During hold off, the oscilloscope will not trigger
until the end of holdoff time. For example, a group of pulse string, it requires that trigger the first pulse string
and the hold off time can be set as width of pulse string, as shown in Figure 6-6.
In HORI MENU use multipurpose knob, jog dial and numeric keyboard to set hold off time.
 Instruments.uni-trend.com
Figure 6-8
Figure 6-9
MSO/UPO2000 Seri es
35 / 137

Advertisement

Table of Contents
loading

Table of Contents