UNI-T MSO/UPO2000 Series Programming Manual page 99

Digital phosphor oscilloscope
Table of Contents

Advertisement

Programming Manual
For example:
:CANFD
:SBUS
Set ID identifier frame data of CAN –FD bus to X00X00X1.
CANFD
:SBUS:
:SBUS:CANFD:DATA:LEN
Command format:
:CANFD
:SBUS
:CANFD
:SBUS
Functional description:
This command is used to set trigger data length of CAN-FD bus of the oscilloscope. It can take value from
1~16.
Return format:
Query returns trigger data length of CAN-FD bus of the oscilloscope, it is integer data.
For example:
:CANFD
:SBUS
:CANFD
:SBUS
:SBUS:CANFD:DATA
Command format:
:CANFD
:SBUS
:CANFD
:SBUS
Functional description:
This command is used to set DATA of CAN-FD bus of the oscilloscope. Binary character string data
presented by parameter is 0, 1 or X. X represents unsure. And its rang is related to the value set by
instruction setting value *8bits :SBUS:CANFD:DATA:LEN, the rang is
0x0~0xFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF.
Return format:
Query returns binary character data.
For example:
:CANFD
:SBUS
:CANFD
:SBUS
:SBUS:CANFD:DATA:OFFSet:CTL
Command format:
:CANFD
:SBUS
:CANFD
:SBUS
Functional description:
This command is used to turn on/off CAN-FD bus DATA byte offset of the oscilloscope.
Return format:
Instruments.uni-trend.com
:ID "X00X00X1"
:ID?
:DATA:LEN <length>
:DATA:LEN?
:DATA:LEN 2
:DATA:LEN?
:DATA <string>
:DATA?
:DATA "X00X00X1"
:DATA?
:OFFSet:CTL
:DATA
{ {1|ON} | {0|OFF} }
:OFFSet:CTL
:DATA
?
Query returns X00X00X1.
Set trigger data length of CAN-FD bus to 2 bytes.
Query returns 2.
Set CAN-FD bus DATA to X00X00X1
Query returns X00X00X1
MSO/UPO2000 Series