UNI-T MSO/UPO2000 Series Programming Manual page 19

Digital phosphor oscilloscope
Table of Contents

Advertisement

Programming Manual
This command is used to set timebase scale of WINDow(<Zoomed> timebase), which is s/div.
Return format:
Query returns < scale> value in scientific notation, unit is s/div.
For example:
:TIM:WIND:SCAL 2
:TIM:WIND:SCAL?
:TIMebase:INDPendent
Command format:
:TIMebase:INDPendent { {1|ON} | {0|OFF} }
:TIMebase:INDPendent?
Functional description:
This command is used to turn on/off timebase independent mode.
Return format:
Query returns 1 or 0, it represents ON or OFF.
For example:
:TIM:INDP ON
:TIM:INDP?
:TIMebase:HOLDoff
Command format:
:TIMebase:
:TIMebase:
Functional description:
This command is used to set trigger holdoff time, which can set the range to 100ns~10s.
Return format:
Query returns the value of trigger holdoff time in scientific notation, unit is s.
For example:
HOLD
:TIM:
HOLD
:TIM:
:TIMebase:SPLit:SCReen
Command format:
:TIMebase:SPLit:SCReen { OFF | CHANnel1| CHANnel2| CHANnel3| CHANnel4|ALL }
:TIMebase:SPLit:SCReen?
Functional description:
This command is used to set the split screen display in main timebase.
OFF: turn off independent display;
CHANnel1| CHANnel2| CHANnel3| CHANnel4: The waveform of channel 1, 2, 3, or 4 displays independently;
ALL: The waveform of all channel displays independently.
Return format:
Query returns {OFF | CHANnel1| CHANnel2| CHANnel3| CHANnel4 | ALL}.
For example:
Instruments.uni-trend.com
HOLDoff
<time>
HOLDoff
?
1s
?
Set the offset of WINDow to 2 s/div.
Query returns 2.000e000.
Turn on timebase independent mode.
Query returns 1.
Set trigger holdoff time to 1s.
Query returns 1.000e000.
MSO/UPO2000 Series
19 / 156