The following steps must be performed to use weighing in the tolerance range:
1. Select function (see chapter 13.1)
2. Set differentiation condition (see chapter 13.2)
3. Set differentiation range (see chapter 13.3)
4. Set discrimination method (see chapter 13.4)
5. Activate / deactivate acoustic signal (see chapter 13.5)
6. Optional: Set relay output (see chapter 13.6)
7. Set tolerance values (see chapter 13.7)

13.1 Selection of weighing function with tolerance range

43
 Press [MENU] key
 Use
[F1]
and
<1 APPLICATIONS>.
 Press [F4] key to go to the application settings.
 Use [F1] and [F2] keys to navigate to <13
COMPARATOR>.
 Press [F4] key to access the settings
 Navigate to <131 ACTIVATE> using the [F1] and
[F2] keys.
 Press the [F4] key
 Use [F1] and [F2] keys to select desired setting
OFF
Disabled
H / L
Lower and upper tolerance limit
HIGH
Upper tolerance limit only
LOW
Lower tolerance limit only
 Press [F4] key to save setting
[F2]
keys
to
TFES_TFEJ-BA-e-2210
navigate
to