Piece Counting; Select The Piece Counting Function - KERN FES 17K-4 Operating Instructions Manual

Precision balance
Table of Contents

Advertisement

10 Piece counting

The piece counting application allows you to count several pieces placed on the
weighing plate.
Before the balance can count parts, it must know the average part weight (i.e. the so-
called reference). Proceed by putting on a certain number of the parts to be counted.
The balance determines the total weight and divides it by the number of parts, the so-
called reference quantity. Counting is then carried out on the basis of the calculated
average piece weight.
As a rule: The higher the reference quantity the higher the counting exactness.
Alternatively, the reference weight can be stored in the scale as a value in advance if
known.
More accurate piece counting is possible by activating Simple SCS. This
function is used with the actual value setting method. For more details on
how to activate and use Simple SCS, refer to Chapter 16.10.
Piece can be counted in two ways:
• Actual value setting method: Weighing the reference piece count (see chapter
10.2)
• Numerical input of piece weight (see chapter 10.3)

10.1 Select the piece counting function

If previous piece counting data is available:
If no previous piece count data is available or value is not accepted:
33
 Select weighing application <COUN> (to select see
chapter 9.5)
Query whether the last piece weight is to be used again:
 Select the desired setting with [F3] and [F4].
NO
Do not import value
YES
Accept value
 <SETTING> display appears and <on 5> flashes.
 Use [F1] and [F2] to select the reference piece
count (see chapter 10.2) or go to numeric entry of
the piece weight (see chapter 10.3))
TFES_TFEJ-BA-e-2210