Connect the power supply unit on the back of the terminal:

7.5.1 Turning On the Power

7.6 Initial Commissioning

The scales must have reached their operating temperature for accurate weighing
results to be obtained from the electronic scales (see Warm-up Time, Chapter 1).
During this warming up time the balance must be connected to the power supply
(mains, accumulator or battery). The accuracy of the balance depends on the local
acceleration of gravity. It is essential that the instructions in the chapter on
Adjustment are followed (see chap.19).

7.7 Connection of peripheral devices

Before connecting or disconnecting of additional devices (printer, PC) to the data
interface, always disconnect the balance from the power supply.
Ensure that the connections of the RS-232C interface and the connection for
peripheral devices do not come into contact with liquids. Cover the connectors on the
device with the protective caps provided when you are not using the connectors.
With your balance, only use accessories and peripheral devices by KERN, as they
are ideally tuned to your balance.
19
 Connect scale to power supply.
 Balance switches to standby mode and an asterisk
(*) is displayed
Mains connection
TFES_TFEJ-BA-e-2210