13.2 Set discrimination condition
The discrimination condition defines whether the evaluation of weight values is
performed only if there are stable weight values or continuously (in case of fluctuating
/ non-stable weight values). The continuous assessment of weight values enables
you to follow in real time on the display during dynamic weighing processes (e.g.
when filling a container) whether your sample is within the tolerance limits.
13.3 Setting the discrimination range
The differentiation range determines the weight value from which the scale starts to
evaluate this value. If the entire range is set, the scale starts at 0 g. If 5 is set, the
evaluation for the weighing systems is carried out according to the following table:
Model
FES 17K-4, FES 33K-4
FES 62K-4D
FEJ 17K-4M, FEJ 33K-4M, FEJ 62K-4DM
13.4 Set discrimination method
TFES_TFEJ-BA-e-2210
Navigate to <132 CONDITION> using the [F1] and
[F2] keys.
Press the [F4] key to change the discrimination
condition.
Use the [F1] and [F2] keys to select the desired
FULL
Always
STBL
Only with steady weighing value
Press [F4] key to confirm selection.
Minimum weight for assessment
0.5 g
0.5 g (up to 6.5 kg) or 5 g (up to 65
kg)
5 g
Use [F1] and [F2] keys to navigate to <133
RANGE>.
Press [F4] key to change the discrimination range
Use the [F1] and [F2] keys to select the desired
5
+5 (e/d) or more
FULL
Total range
Press [F4] key to confirm selection.
Use [F1] and [F2] keys to navigate to <134
METHOD>.
Press [F4] key to change the discrimination method
Use the [F1] and [F2] keys to select the desired
ABSOL
Evaluation of absolute values
RELAT
Evaluation with difference values
Press [F4] key to confirm selection.
44