16.8 Reaction settings

The reaction settings allow you to adjust the scale to the environmental conditions.
The reaction settings influence the stability display of the balance.
The following applies: The higher the value set, the less sensitive the scale is to
environmental influences (e.g. wind or vibration) and is more likely to display a
stable weight value.

16.9 Zero-Tracking

Small weight variations (e.g. due to particles on the weighing plate) can be
automatically tared by zero tracking.
Settings 1, 2 and 4 are not available for the FEJ weighing system.
59
0.5
0.5d
1
1.0d
2
2.0d
4
4.0d
 Press [F4] key to confirm selection.
 Navigate to <2 PERFORMANCE> using the [F1]
and [F2] keys.
 Press [F4] key to access the settings
 Use the [F1] to [F4] keys to navigate to <22
RESPONSE>.
 Press [F4] key to make settings
 Select the desired setting with [F1] and [F2].
1
Very strong sensitivity
2
Strong sensitivity
3
Normal
4
Weak sensitivity
5
Very weak sensitivity (anti-vibration)
 Press [F4] key to confirm selection.
 Navigate to <2 PERFORMANCE> using the [F1]
and [F2] keys.
 Press [F4] key to access the settings