19 Adjustment

As the value of the acceleration due to gravity is not the same at every location on
earth, according to the underlying physical weighing principle, each scale must be
adjusted to the place of installation to the acceleration due to gravity prevailing there
(only if the scale has not already been adjusted to the place of installation at the
factory).
This adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.
Observe stable environmental conditions. A warm up time (see
chapter 1) is required for stabilization.
Ensure that there are no objects on the weighing pan.
Avoid vibration and air flow.
Always carry out adjustment with the standard weighing plate in
place.
To cancel internal adjustment, press the ON/OFF key.
The adjustment record is printed if an optional printer is connected
and the GLP function has been activated.
• The FES weighing system requires external adjustment weights for
adjustment and adjustment.
• The FEJ weighing system has an internal adjustment weight.
The settings for adjustment described in this chapter are always changed in menu
<63 MAINTENANCE>.
73
 Use the [F1] and [F2] keys to navigate to
<6 ADMIN/ADJUST>.
 Press [F4] key to access the settings
 Navigate to <63 MAINTENANCE> using the keys
[F1] and [F2].
 Press [F4] key to access the settings
TFES_TFEJ-BA-e-2210