Input Format 2 - KERN FES 17K-4 Operating Instructions Manual

Precision balance
Table of Contents

Advertisement

• Commands O8 and O9 are used to request data.
• After entering O8 or O9, the scale returns O0.
• Commands O0 to O7 are executed after activation until the scale is
turned off. When the [MENU] key is pressed or the scale is turned on
again, the output settings <413 CONDITION> are reset.
• When OA or OB is entered, the interval output starts. The interval
output stops when these commands are re-entered.
Date and time:
C1
C2
Code (C1)
D
D
44H
D
T
44H
Adjustment / Adjustment test:
C1
C2
Code (C1)
C
1
43H
C
2
43H
C
3
43H
C
4
43H
A00/Exx Format
A00: Normal answer
E01: Incorrect answer
Commands C3 and C4 are only available for the FES weighing system.

21.5.2 Input format 2

Input format:
1
2
3
C1
C2
,
Example for entering an upper tolerance limit (upper limit = 120 g):
 Input: LB,120.0
Example for entering a time for interval output (output every 12 hours, 34 minutes,
and 56 seconds):
 Input: IA,12,34,56 (delimitation by commas).
89
Code (C2)
44H
54H
Code (C2)
31H
32H
33H
34H
Response
4
5
6
7
C3
C3
C3
C3
Significance
Output date
Output time
Significance
Perform internal semi-automatic
adjustment
Perform internal adjustment test
Perform external adjustment
Perform external adjustment test
ACK/NAK Format
ACK: Normal answer
NAK: Incorrect answer
8
9
10
C3
C3
C3
TFES_TFEJ-BA-e-2210
11
12
13
14
C3
C3
C3
CR
15
LF

Advertisement